Agilent 1290 Infinity II 2D-LC Solution User Guide

Contenido

  1. Overview pp. 13-30
    Explains 2D-LC principles, heart-cutting, and comprehensive modes
  2. Overview pp. 31-42
    Describes standard comprehensive 2D-LC operation and ASM
  3. Overview pp. 49-54
    Covers 2D-LC license activation and management
  4. Procedure pp. 55-132
    Installation procedures for valves, decks, and Bio 2D-LC components
  5. Overview pp. 133-144
    Overview of 2D-LC data acquisition in MassHunter Workstation
  6. Procedure pp. 157-228
    Defines pump flow, sampling table, gradient phase, and method parameters
  7. Procedure pp. 229-252
    Instructions for starting system runs and checkout procedures
  8. Procedure pp. 253-306
    2D-LC data analysis and evaluation in MassHunter
  9. Procedure pp. 307-366
    Diagnostic procedures and troubleshooting for common issues
  10. Procedure pp. 367-396
    Cleaning, part replacement, and maintenance procedures
  11. Ordering pp. 397-420
    Available consumables and spare parts for maintenance
  12. Spec pp. 421-426
    Selection guidance for biocompatible LC columns
  13. Overview pp. 427-444
    Theoretical background and practical issues for 2D-LC
  14. Procedure pp. 451-480
    Checkout procedures for multiple heart-cutting and legacy systems
  15. Spec pp. 481-492
    Recommendations on solvent use and stainless steel compatibility
  16. pp. 493-500
    Abbreviation guide for fitting left and right
  17. p. 250 table, 4 rows
  18. p. 465 table, 4 rows
  19. p. 475 table, 4 rows
  20. p. 240 table, 19 rows
  21. p. 241 table, 4 rows
  22. p. 246 table, 4 rows
  23. p. 249 table, 24 rows
  24. p. 450 table, 4 rows
  25. p. 455 table, 4 rows
  26. p. 458 table, 19 rows

Table 38: Recommended conditions in 2D (LC-MS) - SQ MS

ParameterValue
ESI Source ParameterSimilar to the High-end MS parameter
Peak Width0.06 min
SCAN100 – 500 m/z in positive mode
Dwell Time200 ms

Table 70: Recommended conditions in 2D (LC-MS) - SQ MS

ParameterValue
ESI Source ParameterSimilar to the High-end MS parameter
Peak Width0.06 min
SCAN100 – 500 m/z in positive mode
Dwell Time200 ms

Recommended conditions in 2D (LC-MS) - SQ MS

ParameterValue
ESI Source ParameterSimilar to the High-end MS parameter
Peak Width0.06 min
SCAN100 – 500 m/z in positive mode
Dwell Time200 ms

Table 28: Recommended conditions in 2D (LC-MS)

ParameterValue
Ion SourceAtmospheric pressure electrospray (Dual AJS ESI) 6
Ion ModeDual AJS ESI
Ion PolarityPositive
Storage ModeBoth, Centroid preferred
LCMS StreamMS
Acquisition ModeAcquisition Mode MS1 Min Range (m/z) 50 , Max Range (m/z) 500 , Scan Rate (spectra/sec) 3
Instrument ParametersSource Parameters
Gas Temp250 °C
Gas Flow11 L/min
Nebulizer40 psig
SheathGasTemp350 °C
SheathGasFlow12 L/min
Scan Segment1
Scan Source Parameters
Vcap3500 V
Nozzle Voltage300 V
Fragmentor120
Skimmer145
OctopoleRFPeak750

6 For other ion sources than Dual AJS ESI the flow rate may need to be adjusted

Table 29: Recommended conditions in 2D (LC-MS) - SQ MS

ParameterValue
ESI Source ParameterSimilar to the High-end MS parameter
Peak Width0.06 min
SCAN100 – 500 m/z in positive mode
Dwell Time200 ms

Table 33: Recommended conditions in 2D (LC-MS) - SQ MS

ParameterValue
ESI Source ParameterSimilar to the High-end MS parameter
Peak Width0.06 min
SCAN100 – 500 m/z in positive mode
Dwell Time200 ms

Table 37: Recommended conditions in 2D (LC-MS)

ParameterValue
2D Cycle TimeAnalysis 0.25 min (0.35 min if 60 µL sample loop is used), Equilibration 0.10 min
2D Gradient:0 min - 5 % B 0.25 min (0.35 min if 60 µL sample loop is used) - 95 % B
²D Detector (DAD)Diode-array 254 nm, 4 nm BW; Ref 360 nm, 100 nm BW Signal peak width 80 Hz
Stop TimeAs pump/No limit
Ion SourceAtmospheric pressure electrospray (Dual AJS ESI)
Ion ModeDual AJS ESI
Ion polarityPositive
Storage ModeBoth, Centroid preferred
LCMS StreamMS
Acquisition ModeAcquisition Mode MS1 Min Range (m/z) 50 , Max Range (m/z) 500 , Scan Rate (spectra/sec) 3
Instrument ParametersSource Parameters
Gas Temp250 °C
Gas Flow11 L/min
Nebulizer40 psig
SheathGasTemp350 °C
SheathGasFlow12 L/min
Scan Segment1
Scan Source Parameters
Vcap3500 V
Nozzle Voltage300 V
Fragmentor120
Skimmer145
OctopoleRFPeak750
ReferenceMassesRef Mass Enabled Enabled

Table 57: Recommended conditions in 2D (LC-MS) - SQ MS

ParameterValue
ESI Source ParameterSimilar to the High-end MS parameter
Peak Width0.06 min
SCAN100 – 500 m/z in positive mode
Dwell Time200 ms

Table 61: Recommended conditions in 2D (LC-MS) - SQ MS

ParameterValue
ESI Source ParameterSimilar to the High-end MS parameter
Peak Width0.06 min
SCAN100 – 500 m/z in positive mode
Dwell Time200 ms

Table 64: Recommended conditions in 2D (LC-MS)

ParameterValue
Ion SourceAtmospheric pressure electrospray (Dual AJS ESI) 6
Ion ModeDual AJS ESI
Ion PolarityPositive
Storage ModeBoth, Centroid preferred
LCMS StreamMS
Acquisition ModeAcquisition Mode MS1 Min Range (m/z) 50 , Max Range (m/z) 500 , Scan Rate (spectra/sec) 3
Instrument ParametersSource Parameters
Gas Temp250 °C
Gas Flow11 L/min
Nebulizer40 psig
SheathGasTemp350 °C
SheathGasFlow12 L/min
Scan Segment1
Scan Source Parameters
Vcap3500 V
Nozzle Voltage300 V
Fragmentor120
Skimmer145
OctopoleRFPeak750

Engineering Facts

14 facts · every claim linked to a page
Selection

The 2D-LC valve must be hosted in an external valve drive; this constrains the instrument setup.

Determines whether the proposed stack can be used at all; the valve cannot be hosted elsewhere.

Source
“The 2D-LC solution only supports the instrument setup where the 2D-LC valve is hosted in the external valve drive”
Selection

A 1290 Infinity Binary Pump or 1290 Infinity II/III High-Speed Pump is mandatory for the second dimension.

Constrains pump selection for the 2D dimension and prevents using an unsuitable pump.

Source
“In the second dimension, a 1290 Infinity Binary Pump or 1290 Infinity II/III High-Speed Pump is mandatory.”
Selection

MHC and HiRes configurations require additional MHC decks and support both the standard G4236A valve and the ASM G4243A valve head.

Drives hardware bill of materials and stack layout for MHC/HiRes installations.

Source
“2D-LC instruments that are used for Multiple Heart-Cutting or High-Resolution Sampling 2D-LC require additional MHC decks.”
Installation

Different transfer capillaries (6a/6b) must be installed depending on which valve head is used.

Prevents wrong capillary selection and resulting dispersion or connection errors during installation.

Source
“Depending on the valve head that is used, different transfer capillaries (6a/6b) must be installed.”
Installation

A USB hardware dongle is required to activate the 2D-LC solution in addition to software licenses.

Must be ordered and available before commissioning; otherwise the 2D-LC driver cannot be activated.

Source
“Additionally you need a USB hardware dongle to activate the 2D-LC solution.”
Operation

In standard heart-cutting, a second-dimension analysis always finishes, so a later 1D peak may be lost if sampled while 2D is busy.

Explains risk of peak loss and motivates MHC/HiRes or careful timing when using single heart-cutting.

Source
“A started second dimension analysis will always be finished!”
Operation

If all sample loops in the first dimension are occupied, the peak is lost in Multiple Heart-Cutting.

Sets capacity limit of the loop deck and informs method timing and loop count planning.

Source
“If all sample loops in the first dimension are occupied, the peak is lost.”
Operation

Stored cuts are analyzed in reverse order to avoid contamination, and a flush gradient runs before a new parking deck.

Affects sequence logic and carry-over control; engineers must not reorder cuts manually.

Source
“Stored cuts are analysed in backwards order to avoid contamination.”
Operation

In LCxLC the 2D gradient analysis time must be less than or equal to the 1D cut size time.

Defines the modulation constraint; violating it causes incomplete 2D separation and loop conflicts.

Source
“The gradient analysis in the second dimension is less than or equal to the cut size time in the first dimension:”
Performance

ASM dilutes the sample loop content with weak 2D solvent before the 2D column to improve second-dimension separation.

Justifies selecting the ASM valve when 1D solvent strength would distort 2D peaks.

Source
“Active Solvent Modulation (ASM) dilutes the content of the sampling loop (sample and 1D solvent) with weak 2D solvent before it reaches the 2D column”
Performance

The ASM valve has four switching positions; positions 1 and 3 match the standard G4236A valve and positions 2 and 4 add dilution.

Needed to understand valve timing and method programming differences versus the standard valve.

Source
“A full switching cycle of the ASM valve has 4 positions.”
Installation

For a standard 2D-LC with two DADs, the transfer volume is approximately 14 µL.

Sets the expected delay between 1D detection and valve switching; useful for verifying configuration.

Source
“For a standard 2D-LC with two DADs, the transfer volume is approx. 14 µL.”
Operation

If the 2D pump stop time is insufficient to analyze all 2D cuts, an alert appears and the stop time must be extended.

Prevents truncated 2D analysis and lost cuts due to misaligned stop time.

Source
“Most likely the stop time of the 2D pump is not sufficient to analyze all 2D cuts”
Operation

In threshold mode the 2D-LC valve switches when the UV signal rises above the threshold and switches again when it falls below or the cut size elapses.

Defines cut boundaries for peak-triggered methods and helps set threshold and cut size correctly.

Source
“The 2D-LC Valve will switch to the next position when the UV signal falls below this value or the cut size elapse.”