Manual introduction, notices, contents and instructions for use
Overviewpp. 7-8
Terminology and warning symbols used in the manual
Warningpp. 9-12
Safety instructions, combustible mixtures, disposal and service
Specpp. 13-14
Technical data and gauge description
Procedurepp. 15-16
Wiring and electrical connection details
Procedurepp. 17-19
Operating instructions and technical specifications
Procedurepp. 20-22
Gauge tube mounting, use and X-ray limit
Config.p. 23
Available flange configurations and ordering information
Specpp. 24-25
Correction factors for various gases
p. 31
Request for return form and service offices
p. 18table, 9 rows
p. 23table, 7 rows
p. 17table, 3 rows
p. 19table, 9 rows
Table 3 Physical Specifications
Characteristic
Specification
Envelope
Nonex (all models)
Grid
Tungsten “Non-Sag”, 0.025 inch diameter
Filament
Helical tungsten wire
Collector
Tungsten, 0.010 inch diameter
Shield Coating
Platinum, internally connected to filament
Base Leads
Soft nickel, 0.060 inch diameter
Collector Lead
Soft nickel, 0.040 inch diameter
Shipping Weight
3 lbs. (6.6 kg)
Internal Volume
220 cc (not including tubulation)
Table 7 Ordering Information
Part Number
Description
K7360301
572 Ion Gauge, 1 in Nonex tubulation
K7360302
572 Ion Gauge, 1 in Kovar tubulation
K7360303
572 Ion Gauge, 0.75 in Nonex tubulation
K7360304
572 Ion Gauge, 0.75 in Kovar tubulation
K7360305
572 Ion Gauge, 1 in Kovar tubulation, NW25
K7360306
572 Ion Gauge, 1 in Kovar tubulation, NW40
K7360307
572 Ion Gauge, 1 in Kovar tubulation, DN 35 CF-F (2.75 in CFF)
Table 1 Gas Sensitivity Parameters
Gas
Gauge Constant S S = i+/i- x 1/P
Emission Setting for 0.01 Amp/Torr sensitivity (10 uA/micron)
Helium
1.5 per Torr
6.7 mA
Nitrogen
10 per Torr
1.0 mA
Air
10 per Torr
1.0 mA
where: S = Sensitivity, Torr-1; i+ = Ion current, Amps; i- = Grid (electron) current, Amps; P = Partial pressure of the gas, Torr
Table 4 Operating Ratings
Parameter
Rating
Collector
0 VDC (ground)*
Shield
Internally connected to filament
Grid
+180 VDC
Filament
+30 VDC
Filament Voltage
4.0 VAC
Filament Current
3.5 Amps AC (1 mA grid current)
Filament Voltage, absolute maximum
6.0 VAC
Filament Current, absolute maximum
6.0 Amps
Filament Temperature, absolute maximum
1400 °C
* Collector operated at ground potential through electrometer circuit to reduce noise pickup and leakage currents.
Engineering Facts
12 facts · every claim linked to a page
Safety
The gauge is not intrinsically safe and must not be turned on when combustible gases or mixtures are present, because the hot filament and high voltage can act as ignition sources.
Prevents ignition and personal injury when selecting a gauge for systems that may contain combustible gases.
Source
“As with all ionization gauges, this device is not intrinsically safe.”
Installation
The ion gauge controller and vacuum system must be separately grounded to a common ground; a single ground wire between chamber and controller chassis is unsafe.
Improper grounding can expose personnel to lethal voltages up to 1000 V.
Source
“Check that your Ion Gauge Controller and vacuum system are separately grounded to a common ground.”
Installation
Before operating, the system ground must be tested and proven capable of carrying at least 10 A.
Ensures the grounding path can safely handle fault currents and protects operators.
Source
“Before proceeding, test the system ground to be sure that it is complete and capable of supporting at least 10A.”
Selection
The gauge must be paired with a controller that does not limit its broad range or exceed the maximum ratings in the manual.
Prevents loss of measurement range and avoids overstressing the gauge.
Source
“It is important to use a proper gauge controller that does not limit the broad range of the gauge, or exceed any of the maximum ratings noted in this manual.”
Installation
Mount the gauge as close as possible to the measurement location through a large tubulation, preferably 1 inch OD, especially for pressures below 10-8 Torr.
Low conductance causes pronounced pumping and reads lower than actual system pressure.
Source
“In general, the gauge tube should be mounted as close as possible to the desired location through a large tubulation (1" OD recommended).”
Operation
Do not attempt high-temperature baking of the gauge at pressures above 10-5 Torr to avoid oxidation of gauge elements.
Oxidation makes it very difficult to attain ultrahigh vacuum pressure readings.
Source
“High temperature baking of the gauge should not be attempted at pressures above 10-5 to avoid oxidation of the gauge elements”
Maintenance
When degassing, use only the recommended grid voltage and current values to prevent damage to the gauge.
Exceeding degas ratings can sag the grid and permanently damage the gauge.
Source
“In either case, it is important to use the values of grid voltage and current recommended to prevent damage to the gauge (see Table 5).”
Safety
Electron bombardment degassing applies up to 700 VDC to the grid connections, which can be lethal; observe all precautions.
Warns maintenance personnel of a lethal high-voltage hazard during degassing.
Source
“If electron bombardment degassing is employed, high voltages (up to 700 VDC maximum) will be applied to the grid connections; such high voltages can kill.”
Performance
Near the X-ray limit of about 2x10-10 Torr nitrogen equivalent, readings should not be considered unequivocal.
Sets realistic expectations for the lowest reliable pressure measurement.
Source
“Therefore, as the typical value of X-ray limit is approached (2x10-10 Torr nitrogen equivalent), readings should not be considered unequivocal.”
Maintenance
Do not use silicone oil or silicone grease on vacuum equipment; use powder-free butyl or polycarbonate gloves and avoid Alconox on aluminum.
Silicone contamination and Alconox damage can ruin vacuum surfaces and readings.
Source
“Do not use silicone oil or silicone grease.”
Operation
The gauge is composition-sensitive, so readings are only meaningful as equivalent pressure of a known gas, usually nitrogen.
Requires applying gas correction factors when measuring gases other than nitrogen.
Source
“All ionization gauges are composition-sensitive and pressure readings derived from these gauges can only be expressed in terms of equivalent pressure of one gas, usually nitrogen.”
Selection
The 572 has dual tungsten filaments that can be switched externally without disturbing the vacuum system.
Allows filament backup and longer service life without venting the system.
Source
“The 572 ion gauge tube has dual tungsten filaments that can be switched between externally from the gauge controller.”