Analysis of Sulfur Compounds in Petroleum Gases and Natural Gas
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación sobre el análisis de compuestos de azufre en gases de petróleo y gas natural utilizando una columna GC Agilent J&W DB-Sulfur SCD (70 m × 0,53 mm, 4,3 µm, n.º de pieza G3903-63003) con Agilent 7890A GC y 355 SCD; se describen dos conjuntos de condiciones de GC (Condiciones 1 y 2).
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Column length | 70 m p.2 ↗ | DB-Sulfur SCD, p/n G3903-63003 |
| Column inner diameter | 0.53 mm p.2 ↗ | DB-Sulfur SCD, p/n G3903-63003 |
| Column film thickness | 4.3 µm p.2 ↗ | DB-Sulfur SCD, p/n G3903-63003 |
| Carrier gas pressure | 8.6 psi p.2 ↗ | Helium, constant pressure mode |
| Inlet temperature | 240 °C p.2 ↗ | Conditions 1, inert flow path split/splitless inlet |
| Split ratio | 10:1 p.2 ↗ | Conditions 1 |
| Oven initial temperature | 35 °C p.2 ↗ | held for 2 min |
| Oven ramp 1 | 35 to 80 at 10 °C/min p.2 ↗ | |
| Oven hold at 80 °C | 6 min p.2 ↗ | |
| Oven ramp 2 | 80 to 220 at 15 °C/min p.2 ↗ | |
| Oven final hold | 220 for 10 °C, min p.2 ↗ | |
| Sample loop volume | 1 mL p.2 ↗ | |
| Burner temperature | 800 °C p.2 ↗ | SCD conditions |
| Burner vacuum | 350 Torr p.2 ↗ | SCD conditions |
| Reaction cell vacuum | 5 Torr p.2 ↗ | SCD conditions |
| Hydrogen flow | 40 mL/min p.2 ↗ | SCD conditions |
| Air flow | 60 mL/min p.2 ↗ | SCD conditions |
Operating Limits
- Maximum oven temperature: 220 °C p.2 ↗
- Burner vacuum: 350 Torr p.2 ↗
- Reaction cell vacuum: 5 Torr p.2 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | Aplica a | Note |
|---|---|---|---|---|
| Agilent J&W DB-Sulfur SCD, 70 m × 0.53 mm, 4.3 µm p.2 ↗ | G3903-63003 | GC column | ||
| Deactivated fused silica tubing, 1 m × 0.18 mm p.2 ↗ | 160-2615-1 | Tubing | ||
| Deactivated Ultimate union p.2 ↗ | G3182-60580 | Union | ||
| Ultra Inert liner p.2 ↗ | 5190-2295 | Liner |