AGILENT GAS CHROMATOGRAPHY AND
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación 5991-6552EN del detector de quimioluminiscencia de azufre (SCD) Agilent 8355; abarca las condiciones del método GC/SCD y los límites de detección para compuestos de azufre en combustibles gaseosos
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Column | Agilent DB-Sulfur, 320 µm × 60 m, 4.2 µm p.2 ↗ | |
| Column part number | G3903-63001 p.2 ↗ | |
| Sample introduction | 2-Valve system p.2 ↗ | |
| Carrier gas | Constant pressure at 14.5 psi p.2 ↗ | |
| Oven temperature program | 30 °C (1.5 min) then 15 °C/min to 250 °C (3 min) °C p.2 ↗ | |
| SCD base temperature | 250 °C p.2 ↗ | |
| SCD furnace temperature | 800 °C p.2 ↗ | |
| Air flow (oxidation) | 60 mL/min p.2 ↗ | |
| H2 flow (oxidation/lower) | 38 mL/min p.2 ↗ | |
| H2 flow (upper) | 8 mL/min p.2 ↗ | |
| O2 flow (ozone generator) | 40 mL/min p.2 ↗ | |
| Detection limit, Hydrogen sulfide | 0.096 pg/sec p.2 ↗ | 2.3 ppm standard |
| Detection limit, Carbonyl sulfide | 0.20 pg/sec p.2 ↗ | 2.3 ppm standard |
| Detection limit, Methyl mercaptan | 0.49 pg/sec p.2 ↗ | 2.3 ppm standard |
| Detection limit, Ethyl mercaptan | 1.04 pg/sec p.2 ↗ | 2.3 ppm standard |
| Detection limit, Dimethyl sulfide | 0.20 pg/sec p.2 ↗ | 2.3 ppm standard |
| Detection limit, Carbon disulfide | 0.094 pg/sec p.2 ↗ | 2.3 ppm standard |
| Detection limit, 2-Propanethiol | 1.42 pg/sec p.2 ↗ | 2.3 ppm standard |
| Detection limit, tert-Butyl mercaptan | 1.62 pg/sec p.2 ↗ | 2.3 ppm standard |
| Detection limit, 1-Propanethiol | 4.3* pg/sec p.2 ↗ | 2.3 ppm standard |
| Detection limit, Thiophene | 0.21 pg/sec p.2 ↗ | 2.3 ppm standard |
Operating Limits
- SCD furnace temperature: 800 °C p.2 ↗
- SCD base temperature: 250 °C p.2 ↗
- Oven maximum temperature: 250 °C p.2 ↗