Cannabinoid Profiling and Quantitation in Hemp Extracts using
Technical Reference Summary
Covered Models & Document Scope
Resumen de aplicación sobre el perfilado y la cuantificación de cannabinoides en extractos de cáñamo mediante el Agilent 1290 Infinity II UHPLC con Agilent 6230B LC-TOF MS; comparación con el 1290 Infinity II UHPLC con DAD. No se indican variantes de modelo de instrumento ni códigos de pedido.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Column | Agilent ZORBAX Bonus RP. 2.1 × 50 mm, 1. p.2 ↗ | |
| Stop time | 12.5 minutes p.2 ↗ | |
| Post time | 1.0 minutes p.2 ↗ | |
| Overall run time | 20.0 minutes p.2 ↗ | including re-equilibration |
| Nebulizer gas pressure | 40 psi p.2 ↗ | |
| Capillary voltage | 4,000 V p.2 ↗ | |
| Fragmentor | 175 V p.2 ↗ | |
| Skimmer | 65 V p.2 ↗ | |
| Ref mass enabled | Yes p.2 ↗ | |
| Detection limit | 1.0 ng/mL p.1 ↗ | as low as |
| Linear regression coefficient | 0.999 p.1 ↗ | or better |
| Calibration range | 10 to 1,000 µg/mL p.1 ↗ | |
| THC concentration in hemp products | <0.3 % p.1 ↗ |
Operating Limits
- Detection limit (lowest): 1.0 ng/mL p.1 ↗
- Calibration range: 10 µg/mL to 1,000 µg/mL p.1 ↗
- THC content in hemp products: <0.3 % p.1 ↗