Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que cubre la configuración del sistema Agilent 1260 Infinity II Bio-inert LC (bomba Bio-inert G5654A, multisampler Bio-inert G5668A con opción de enfriador de muestras #100, termostato multicolumna G7116A con opción de intercambiador de calor bio-inert #019, detector de matriz de diodos WR G7115A con opción de celda de flujo bio-inert #028) utilizado con columnas AdvanceBio SEC 300 Å, 7,8 × 300 mm, 2,7 µm (n/p PL1180-5301) y Bio SEC-3, 300 Å, 7,8 × 300 mm, 3 µm (n/p 5190-2511); software OpenLAB CDS versión 2.1.
Technical Data
| Parameter |
Value |
Condición |
| Column dimensions |
7.8 × 300 mm
p.2 ↗ |
AdvanceBio SEC 300 Å, 2.7 µm (p/n PL1180-5301) |
| Column dimensions |
7.8 × 300 mm
p.2 ↗ |
Bio SEC-3, 300 Å, 3 µm (p/n 5190-2511) |
| Column particle size |
2.7 µm
p.2 ↗ |
AdvanceBio SEC 300 Å (p/n PL1180-5301) |
| Column pore size |
300 Å
p.2 ↗ |
AdvanceBio SEC (p/n PL1180-5301) |
| Column pore size |
300 Å
p.2 ↗ |
Bio SEC-3 (p/n 5190-2511) |
| Injection volume |
10 µL
p.2 ↗ |
|
| Needle wash mode |
Standard Wash
p.2 ↗ |
|
| Salt concentration tolerance |
2 M NaCl
p.2 ↗ |
Bio-inert flowpath |
| Urea concentration tolerance |
up to 8 M urea
p.2 ↗ |
Bio-inert flowpath |
| NaOH concentration tolerance |
0.5 M NaOH
p.2 ↗ |
Bio-inert flowpath |
| HCl concentration tolerance |
0.5 M HCl
p.2 ↗ |
Bio-inert flowpath |
| Software version |
2.1
p.2 ↗ |
OpenLAB CDS |
Operating Limits
- Maximum salt concentration (NaCl): 2 M
p.2 ↗
- Maximum urea concentration: 8 M
p.2 ↗
- Maximum NaOH concentration: 0.5 M
p.2 ↗
- Maximum HCl concentration: 0.5 M
p.2 ↗
Type Code Decoder
| Position |
Token |
Meaning |
| field 1
p.2 ↗ |
G5654A |
Agilent 1260 Infinity II Bio-inert Pump |
| field 2
p.2 ↗ |
G5668A |
Agilent 1260 Infinity II Bio-inert Multisampler |
| field 3
p.2 ↗ |
G7116A |
Agilent 1260 Infinity II Multicolumn Thermostat |
| field 4
p.2 ↗ |
G7115A |
Agilent 1260 Infinity II Diode Array Detector WR |
Ordering Data
| Ordering Code |
Part No. |
Variant / Code |
Aplica a |
Note |
| Agilent 1260 Infinity II Bio-inert Pump (G5654A)
p.2 ↗ |
G5654A |
|
|
System module |
| Agilent 1260 Infinity II Bio-inert Multisampler (G5668A) with sample cooler (option #100)
p.2 ↗ |
G5668A |
#100 |
with sample cooler option |
System module |
| Agilent 1260 Infinity II Multicolumn Thermostat (G7116A) with bio-inert heat exchanger (option #019)
p.2 ↗ |
G7116A |
#019 |
with bio-inert heat exchanger option |
System module |
| Agilent 1260 Infinity II Diode Array Detector WR (G7115A) with bio-inert flow cell (option #028)
p.2 ↗ |
G7115A |
#028 |
with bio-inert flow cell option |
System module |
| Agilent AdvanceBio SEC 300 Å, 7.8 × 300 mm, 2.7 µm
p.2 ↗ |
PL1180-5301 |
|
|
SEC column |
| Agilent Bio SEC-3, 300 Å, 7.8 × 300 mm, 3 µm
p.2 ↗ |
5190-2511 |
|
|
SEC column |