Technical Reference Summary
Covered Models & Document Scope
Guía del usuario para columnas de HPLC Agilent Altura con tecnología Ultra Inert, Agilent AdvanceBio y columnas de fase inversa Agilent BioHPLC. Columnas cubiertas: Altura HPH-C18 1.9 µm, Altura Oligo HPH-C18 2.7 µm, Altura Peptide Plus, Altura PLRP-S, Altura ZORBAX Eclipse Plus C18, AdvanceBio Peptide Mapping, AdvanceBio Peptide Plus, AdvanceBio Oligonucleotide, AdvanceBio Amino Acid Analysis, AdvanceBio RP-mAb, PLRP-S, ZORBAX 300, Poroshell 300.
Technical Data
| Parameter |
Value |
Condición |
| Maximum system pressure |
Up to 400 bar
p.2 ↗ |
InfinityLab Quick Turn fitting (finger-tight) |
| Recommended flow rate |
0.1 mL/min
p.2 ↗ |
Column id 2.1 mm |
| Recommended temperature |
60 °C
p.2 ↗ |
Column id 2.1 mm |
| Recommended temperature |
60 °C
p.2 ↗ |
Column id 3.0 mm |
| Recommended temperature |
60 °C
p.2 ↗ |
Column id 4.6 mm |
| Column volume |
50
p.2 ↗ |
Column id 2.1 mm |
| Column volume |
50
p.2 ↗ |
Column id 3.0 mm |
| Column volume |
50
p.2 ↗ |
Column id 4.6 mm |
| Particle size |
1.9 µm
p.3 ↗ |
Altura HPH-C18 |
| Maximum pressure |
1200 bar
p.3 ↗ |
Altura HPH-C18 |
| pH range |
1.0 to 8.0
p.3 ↗ |
AdvanceBio RP-mAb |
| Maximum temperature |
90 °C
p.3 ↗ |
AdvanceBio RP-mAb |
Operating Limits
- Maximum system pressure with Quick Turn fitting (finger-tight): 400 bar
p.2 ↗
- Maximum system pressure with Quick Turn fitting (with mounting tool): 800 bar
p.2 ↗
- Maximum system pressure with Quick Connect fitting: 1300 bar
p.2 ↗
- Maximum temperature for AdvanceBio RP-mAb: 90 °C
p.3 ↗
- Maximum temperature for Altura PLRP-S and PLRP-S: 200 °C
p.3 ↗
- Maximum shipping solvent storage temperature: 40 °C
p.4 ↗
Ordering Data
| Ordering Code |
Part No. |
Variant / Code |
Aplica a |
Note |
| InfinityLab Quick Connect assembly
p.2 ↗ |
5067-5961 |
|
|
Fitting assembly |
| InfinityLab Quick Turn fitting
p.2 ↗ |
5067-5966 |
|
|
Finger-tight or with mounting tool |
| Mounting Tool
p.2 ↗ |
5043-0915 |
|
|
For Quick Turn fitting |
| InfinityLab Quick Connect fitting
p.2 ↗ |
5067-5965 |
|
|
For up to 1300 bar |