8890 GC site preparation checklist D0007040 en agilent
Technical Reference Summary
Covered Models & Document Scope
Lista de verificación de preparación del sitio para el cromatógrafo de gases Agilent 8890A/B (documento D0007040, Rev. 3.0). Abarca el GC 8890/8890B con rampa de horno estándar o rápida, configuraciones de 120 VAC para América o de 220/230/240 VAC, además del PC del sistema de datos y las opciones de cable de alimentación.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Height | 50 cm p.9 ↗ | 8890 Series GC |
| Width | 54 cm p.9 ↗ | 8890 Series GC |
| Depth | 54 cm p.9 ↗ | 8890 Series GC |
| Weight | 50 kg p.9 ↗ | 8890 Series GC |
| Operating temperature range | 15 to 35 °C p.9 ↗ | Standard oven ramp |
| Operating temperature range | 15 to 35 °C p.9 ↗ | Fast oven ramp (option 002) |
| Operating humidity range | 5 to 95 % p.9 ↗ | Operation |
| Maximum altitude | 4,615 m p.9 ↗ | Operation |
| Storage humidity range | 5 to 95 % p.9 ↗ | Storage |
| Line voltage | 120 VAC p.11 ↗ | 8890/8890B GC, Standard oven, Americas |
| Frequency | 50/60 ± 5% Hz p.11 ↗ | 8890/8890B GC |
| Power rating | 2350 W p.11 ↗ | 8890/8890B GC, Standard oven |
| Current rating | 19.6 amps p.11 ↗ | 8890/8890B GC, Standard oven, 120 VAC |
| Power outlet current | 20 Amp p.11 ↗ | 8890/8890B GC, Standard oven, Americas 120 VAC |
Operating Limits
- Operating temperature range: 15 to 35 °C p.9 ↗
- Operating humidity range: 5 to 95% p.9 ↗
- Maximum altitude: 4,615 m p.9 ↗
- Storage temperature range: –40 to 70 °C p.9 ↗
- Storage humidity range: 5 to 95% p.9 ↗
- Mains frequency tolerance: 50/60 ± 5% Hz p.11 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | Aplica a | Note |
|---|---|---|---|---|
| 8121-3119 p.12 ↗ | 8121-3119 | Argentina, 220 V, 16 A, C19, IRAM 2073 | Power cord | |
| 8121-3115 p.12 ↗ | 8121-3115 | Australia/New Zealand, 230 V, 10 A, C13, AS 3112 | Power cord | |
| 8120-8619 p.12 ↗ | 8120-8619 | Australia/New Zealand, 230 V, 15 A, C19, AS 3112 | Power cord | |
| 8121-3321 p.12 ↗ | 8121-3321 | Brazil, 120/220 V, 16 A, C19, NBR 14136 BR/3/20 | Power cord | |
| 8121-1084 p.12 ↗ | 8121-1084 | Chile/Italy, 230/220 V, 16 A, C19, CEI 23-50 | Power cord | |
| 8121-1766 p.12 ↗ | 8121-1766 | China, 220 V, 16 A, C19, GB 1002 GB 2099 | Power cord | |
| 8121-0723 p.12 ↗ | 8121-0723 | China, 220 V, 10 A, C13, GB 1002 GB 2099 | Power cord | |
| 8120-8622 p.12 ↗ | 8120-8622 | Denmark, 230 V, 16 A, C19, IEC60309 | Power cord | |
| 8121-3112 p.12 ↗ | 8121-3112 | Denmark, 230 V, 10 A, C13, SB 107-2-D1 DK2-5a | Power cord | |
| 8120-4416 p.12 ↗ | 8120-4416 | Switzerland, 230 V, 10 A, C13, SEV 1011 | Power cord |