Bacterial Artificial Chromosome (BAC) Analysis

Contenido

  1. Overview p. 2
    Background on BAC clones and analysis needs
  2. Procedure p. 3
    Workflow for BAC DNA sample preparation and analysis
  3. p. 4
    Experimental results and interpretation of BAC data
  4. Spec p. 5
    Sensitivity performance data for the Femto Pulse system
  5. p. 6
    Summary of findings and recommendations
  6. p. 2 table, 6 rows
  7. p. 4 table, 4 rows
  8. p. 5 table, 5 rows

Table 1. Comparison of PFGE and the Agilent Femto Pulse system.

AttributePFGE*Agilent Femto Pulse system
Samples per Run1 to 141 to 11
Sample AmountNanogramsPicograms
Separation Time10 to 24 hours3 hours
Total Time (Inc. Gel Prep., Separation, and Staining)12 to 26 hours3 hours
Size RangeUp to 10 Mb75 bp to 165 kb
Result FormatAgarose gel image; requires manual annotation and documentationDigital gel image and electropherogram; automated sizing analysis

*Attributes based on Bio-Rad Chef Mapper XA system

Table 2. Comparison of the expected BAC and vector sizes reported by the Agilent Femto Pulse system with the Agilent 165 kb BAC kit. (Note: the largest fragment splits into two partially resolved peaks. For sizing, the peaks were analyzed together using the smear analysis function in Agilent ProSize data analysis software.)

BAC Expected Size (bp)BAC D1 Femto Pulse Size (bp)Vector Expected Size (bp)Vector D2 Femto Pulse Size (bp)
98,18693,008101,24993,968
32,25731,08632,25730,390
6,1657,3072,6552,692
2,6553,125

Table 3. Comparison of the expected BAC and vector sizes to the sizes reported by the Agilent Femto Pulse with the Agilent 165 kb BAC kit.

BAC Expected Size (bp)BAC B5 Femto Pulse Size (bp) (Pass)Vector (B6, B7) Expected Size (bp)Vector B6 Femto Pulse Size (bp) (Fail)Vector B7 Femto Pulse Size (bp) (Fail)
47,53747,65351,53445,62246,186
11,47212,19411,47211,63311,531
4574556,7154,876325
3101,695457453449
325310308

Engineering Facts

12 facts · every claim linked to a page
Selection

The Femto Pulse with the 165 kb BAC kit resolves fragments up to 165 kb, so it can replace PFGE for BAC digest sizing within that range.

Defines the upper sizing limit that determines whether this instrument fits the BAC fragment sizes in the workflow.

Source
“such as digested BAC samples up to 165 kb in less than 3 hours”
Performance

Separation takes about 3 hours versus 10 to 24 hours for PFGE, removing the overnight bottleneck.

Run time drives throughput planning and lab scheduling when choosing between PFGE and capillary electrophoresis.

Source
“Separation Time 10 to 24 hours 3 hours”
Selection

The system needs only picogram-level sample input versus nanograms for PFGE, which matters when BAC material is scarce.

Sample consumption is a key selection criterion for precious BAC and vector digests.

Source
“Sample Amount Nanograms Picograms”
Installation

The instrument runs up to 11 samples per run, so batch capacity must be planned accordingly.

Throughput per run determines how many digests can be batched and how the lab schedules QC.

Source
“Samples per Run 1 to 14 1 to 11”
Operation

Salts in the digest must be diluted out per Appendix A before loading, otherwise mobility and sizing may be affected.

Sample matrix handling is a mandatory prep step to obtain reliable sizing results.

Source
“Salts present in the samples were diluted out following the additional sample preparation instructions in Appendix A of the quick guide”
Operation

The digestion reaction is diluted 10X in Dilution Buffer 0.25x TE, then 2 µL is mixed with 18 µL gDNA Diluent Marker using a wide bore pipette.

Exact dilution and pipetting steps must be followed to avoid sizing errors and sample loss.

Source
“the digestion reaction was diluted 10X using Dilution Buffer 0.25x TE”
Operation

The 165 kb BAC Ladder must be prepared for a 10X sample predilution and run in parallel with samples to compensate for matrix effects on mobility.

Correct ladder preparation is required for accurate, matrix-matched sizing.

Source
“the 165 kb BAC Ladder working solution was prepared according to the specifications for a 10X sample predilution”
Performance

The largest fragment can split into two partially resolved peaks, which must be analyzed together using smear analysis in ProSize.

Data analysis settings affect reported sizes and must be configured for correct QC calls.

Source
“the largest fragment splits into two partially resolved peaks. For sizing, the peaks were analyzed together using the smear analysis function”
Performance

Small fragments below 500 bp are lost on PFGE but detected by the Femto Pulse, giving higher sensitivity in the low size range.

Detection of small fragments is decisive for confirming recombination and avoiding false passes.

Source
“The two fragments with expected sizes below 500 bp cannot be visualized on the PFGE gel image”
Operation

PFGE comparison runs used a 1% PFGE-grade agarose gel and a 10.5-hour method for 2 to 200 kb fragments.

Defines the reference PFGE method parameters needed for a valid side-by-side comparison.

Source
“a method to allow for separation of fragments between 2 to 200 kb for 10.5 hours”
Maintenance

PFGE gels require post-staining with 10X Gel Red and UV imaging at 365 nm, adding staining and destaining time to the workflow.

Extra staining and imaging steps explain the longer total time and consumable burden of PFGE.

Source
“The agarose gel was post-stained with a 10X Gel Red solution”
Cumplimiento

The system is for research use only and not for diagnostic procedures, which constrains its deployment.

Regulatory labeling determines whether the instrument can be used in clinical or diagnostic settings.

Source
“For Research Use Only. Not for use in diagnostic procedures.”