Induztr
Resumen técnico Visor de documentos Modelos relacionados

Forced Degradation Studies of Synthetic Oligonucleotide

Technical Reference Summary

Covered Models & Document Scope

Nota de aplicación sobre estudios de degradación forzada de Fomivirsen utilizando un sistema bio LC Agilent 1290 Infinity II acoplado a un LC/Q-TOF Agilent 6545XT AdvanceBio; abarca condiciones de estrés térmico (80 °C, 4/8/24 h) y estrés oxidativo (0.30% y 3.00% H2O2, 2/4 h).

Technical Data

Parameter Value Condición
Solvent A composition 15 mM TEA and 100 mM HFIP in water p.2 ↗
Sample concentration 1 mg/mL p.2 ↗
Heat stress incubation temperature 80 °C p.2 ↗
Heat stress incubation time points 4, 8, and 24 hours p.2 ↗
Oxidative stress H2O2 concentration 0.30 and 3.00 % p.2 ↗
Oxidative stress duration 2 and 4 hours p.2 ↗
Source Dual AJS p.3 ↗
Gas Flow 12 L/min p.3 ↗
Nebulizer 35 psi p.3 ↗
Sheath Gas Flow 12 L/min p.3 ↗
Skimmer 65 V p.3 ↗
Tune Mode Extended Dynamic Range (2 GHz) p.3 ↗
Acquisition Mode MS Scan p.3 ↗
Mass Range m/z 300 to 3,200 p.3 ↗
MS Scan Range m/z 400 to 3,000 p.3 ↗
Software version 1.5 p.3 ↗ Agilent MassHunter BioConfirm Software
Match Tolerance Mass ±10 ppm p.3 ↗

Operating Limits

  • Sample storage temperature: –80 °C p.2 ↗
  • Heat stress incubation temperature: 80 °C p.2 ↗
  • Oxidative stress H2O2 concentration: 0.30 and 3.00% p.2 ↗