Hydrogen Carrier Gas for Analyzing Pesticides in Pigmented Foods with GC/MS/MS

Technical Reference Summary

Covered Models & Document Scope

Nota de aplicación sobre el análisis de pesticidas por GC/TQ con gas portador de hidrógeno, utilizando sistemas Agilent 8890 GC, 7000E y 7010C GC/TQ, fuentes de ionización electrónica HydroInert y HES, columna HP-5ms UI de 20 m × 0,18 mm × 0,18 µm, y 203 pesticidas objetivo en matriz de espinaca pigmentada

Technical Data

Parameter Value Condición
Column length 20 m p.4 ↗ Agilent HP-5ms UI (p/n 19091S-577UI)
Column diameter 0.18 mm p.4 ↗ Agilent HP-5ms UI
Film thickness 0.18 µm p.4 ↗ Agilent HP-5ms UI
Column flow 1.0 mL/min p.4 ↗ constant flow, nominal before retention time locking
Inlet temperature 60 °C for 0.06 min, then to 280 °C at 60 °C p.4 ↗ Multimode Inlet (MMI), solvent vent mode
Initial oven temperature 60 °C p.4 ↗ oven program
Oven ramp rate 2 10 °C/min p.4 ↗ oven program
Total run time 20 min p.4 ↗ 10-minute analysis with backflush
Purge flow to split vent 60 mL/min p.4 ↗ at 2.56 min
Septum purge flow 3 mL/min p.4 ↗ switched mode
Vent flow 100 mL/min p.4 ↗ solvent vent mode
Vent pressure 5 psi p.4 ↗ until 0.06 min
Cryo use temperature 200 °C p.4 ↗ cryo on (air)
Total MRMs 614 p.4 ↗ dMRM mode
Minimum dwell time 3 ms p.4 ↗ dMRM mode
EM voltage gain mode 10 p.4 ↗ MSD
Vacuum pump Performance turbo p.4 ↗ Agilent 8890 GC/TQ system
Quantitation range 0.1 to 1,000 ppb p.14 ↗ 7010C GC/TQ with HES, quadratic fit
Default MRL 10 ppb p.2 ↗ spinach matrix
Target pesticides quantified at or below MRL over 90% of 203 % p.2 ↗ HydroInert and HES sources
RSE below 20% over 94% of 203 % p.2 ↗ SANTE/11312/2021 guidelines
Analyte response enhancement 10-fold on average fold p.1 ↗ optimized injection conditions vs helium injection conditions with hydrogen
MDL for reactive compounds sub-ppb ppb p.2 ↗ HES and HydroInert with hydrogen
Lowest detectable level 0.5 ppb p.11 ↗ HydroInert with H2

Operating Limits

  • Maximum oven temperature: 310 °C p.8 ↗
  • Minimum calibration concentration: 0.1 ppb p.14 ↗
  • Maximum calibration concentration: 1,000 ppb p.14 ↗
  • Default MRL: 10 ppb p.2 ↗