Analysis of Sulfur Compounds in High-Purity Hydrogen Using an Agilent GC/SCD
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que cubre el análisis de compuestos de azufre en hidrógeno utilizando un cromatógrafo de gases Agilent 8890 configurado con un detector de quimioluminiscencia de azufre (SCD) Agilent 8355; no se indican variantes de modelo/tamaño de producto ni códigos de pedido.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Sample introduction | six-port gas sampling valve connected directly to an inert volatiles interface (VI) p.2 ↗ | 8890 GC with 8355 SCD |
| Sample loop volume | 2 mL p.2 ↗ | 8890 GC with 8355 SCD |
| Split ratio | 10:1 p.2 ↗ | 8890 GC with 8355 SCD |
| Data acquisition rate | 5 Hz p.3 ↗ | 0.04 minutes |
| Calibration range | 15 to 200 ppb p.5 ↗ | sulfur compounds |
| Calibration level | 15 ppb p.5 ↗ | low level |
| Calibration mix concentration | 0.992 µmol/mol, ppm p.3 ↗ | Hydrogen sulfide (H2S) |
| Retention time | 3.976 min p.5 ↗ | Hydrogen sulfide |
| Correlation coefficient (R2) | 0.9987 p.5 ↗ | Hydrogen sulfide |
Operating Limits
- Calibration range: 15 to 200 ppb p.5 ↗
- Data acquisition rate: 5 Hz / 0.04 minutes p.3 ↗
- Sample loop volume: 2 mL p.2 ↗
- Split ratio: 10:1 p.2 ↗