An End-To-End Workflow for Quantitative Screening of Multiclass,
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que cubre un flujo de trabajo LC/MS/MS de extremo a extremo para el cribado cuantitativo de 210 fármacos veterinarios multiclase y multirresiduo en carne (pollo, cerdo, vacuno) utilizando LC/MS de triple cuadrupolo Agilent 6470 y 6495C con LC Agilent 1290 Infinity II, limpieza Captiva EMR—Lipid y columnas InfinityLab Poroshell 120 EC-C18.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Number of target analytes | 210 targets p.1 ↗ | multiclass veterinary drugs |
| Matrices evaluated | chicken, pork, beef p.1 ↗ | muscle |
| LC/MS models | 6470 and 6495C p.1 ↗ | triple quadrupole LC/MS |
| Chromatographic method time | 13 min p.1 ↗ | single method |
| Column | InfinityLab Poroshell 120 EC-C18 p.1 ↗ | |
| Spike/calibration range | 0.1 to 100 μg/L p.1 ↗ | matrix-matched spike samples |
| Calibration curve linearity R2 | >0.99 p.1 ↗ | LOQ to 100 μg/L |
| Instrument method accuracy | 73 to 113 % p.1 ↗ | |
| Retention time %RSD | ≤0.28 % p.1 ↗ | |
| Analyte response %RSD | ≤19 % p.1 ↗ | |
| QC fortification levels | 1, 10, 25 μg/kg p.1 ↗ | LQC, MQC, HQC in meat |
| Recovery repeatability limit | 20 % p.1 ↗ | intrabatch technical replicates |
| Recovery reproducibility limit | 32 % p.1 ↗ | interbatch technical replicates |
| Injection loop volume | 20 µL p.4 ↗ | 1290 Infinity II high-speed pump (G4220A) |
| Mobile phase A | water with 4.5 mM ammonium formate, 0.5 mM ammonium fluoride, 0.1% formic acid p.4 ↗ | |
| Mobile phase B | 50/50 ACN/MeOH with 4.5 mM ammonium formate, 0.5 mM ammonium fluoride p.4 ↗ | |
| Typical chromatographic peak width | 8 to 12 s p.5 ↗ | |
| MRM transitions per compound | 2 or 3 transitions p.5 ↗ | target-specific |
| Targets meeting recovery criteria | 210 targets p.8 ↗ | all targets |
| Targets with low ion suppression | >93 % p.8 ↗ | matrix effect comparison |
| Targets with medium ion suppression | 1 % p.8 ↗ | matrix effect 25 to <50% |
| Recovery of targets in chicken | >97 % p.8 ↗ | 10 μg/kg MQC |
| Recovery range for some targets | 30 to 60 % p.7 ↗ | malachite green, narasin, nicotine |
| Targets with no MRL established | 42 targets p.4 ↗ | monitoring category |
| Targets specific to AOAC recommended list | 86 targets p.4 ↗ | out of 168 targets |
Operating Limits
- Calibration range: LOQ to 100 μg/L p.1 ↗
- Spike sample range: 0.1 to 100 μg/L p.1 ↗
- QC fortification levels: 1, 10, 25 μg/kg p.1 ↗
- Recovery repeatability acceptance limit: 20% p.1 ↗
- Recovery reproducibility acceptance limit: 32% p.1 ↗
- Retention time %RSD: ≤0.28% p.1 ↗