Dioxinanalyse in Lebens- und Futtermitteln mit dem
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación (publicación de Agilent 5991-9399DEE) sobre el análisis de dioxinas en alimentos y piensos utilizando el sistema Agilent Intuvo 9000/7010 GC Triple Quadrupole; abarca dos métodos de GC: un método de 52,5 minutos y un método acelerado de 31,12 minutos.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Column flow rate | 1,6 ml/min p.2 ↗ | 52.5-minute method |
| Column flow rate | 1,6 ml/min p.2 ↗ | 31.12-minute method |
| Oven temperature program | 20 °C/min bis 220 °C (15 Minuten) p.2 ↗ | 52.5-minute method |
| Guard Chip temperature | 310 °C p.2 ↗ | 52.5-minute method |
| Guard Chip temperature | 310 °C p.2 ↗ | 31.12-minute method |
| Bus temperature | 325 °C p.2 ↗ | 52.5-minute method |
| Bus temperature | 325 °C p.2 ↗ | 31.12-minute method |
| Source temperature | 280 °C p.2 ↗ | 52.5-minute method |
| Source temperature | 280 °C p.2 ↗ | 31.12-minute method |
| Retention time 2378-TCDF | 26,64 min p.4 ↗ | 31.12-minute method |
| Retention time 2378-TCDD | 27,835 min p.4 ↗ | 31.12-minute method |
| Retention time OCDD | 48,83 min p.4 ↗ | 31.12-minute method |
| Retention time OCDF | 48,995 min p.4 ↗ | 31.12-minute method |
| MDL 2378-TCDF | 3,8184 fg/µl p.4 ↗ | 31.12-minute method |
| MDL 2378-TCDD | 5,7684 fg/µl p.4 ↗ | 31.12-minute method |
| LOQ 2378-TCDF | 13,5334 fg/µl p.4 ↗ | 31.12-minute method |
| LOQ 2378-TCDD | 20,445 fg/µl p.4 ↗ | 31.12-minute method |
| LOD 2378-TCDF | 4,06 fg/µl p.4 ↗ | 31.12-minute method |
| LOD 2378-TCDD | 6,1335 fg/µl p.4 ↗ | 31.12-minute method |
Operating Limits
- Maximum oven temperature (52.5-minute method): 330 °C p.2 ↗
- Maximum oven temperature (31.12-minute method): 310 °C p.2 ↗
- Guard Chip temperature: 310 °C p.2 ↗
- Bus temperature: 325 °C p.2 ↗
- Source temperature: 280 °C p.2 ↗