Data Sheet Model ED S 56

Technical Reference Summary

Covered Models & Document Scope

Cámara de secado y calentamiento modelo ED-S 56 con convección natural, Solid.Line. Dos variantes: EDS056-230V (Art.-No. 9090-0014, 230 V 1~ 50/60 Hz, enchufe CEE 7/7) y EDS056UL-120V (Art.-No. 9090-0015, 120 V 1~ 60 Hz, enchufe NEMA 5-15).

Technical Data

Parameter Value Condición
Temperature range +7 °C above ambient temperature to 250 °C p.2 ↗ EDS056-230V
Temperature uniformity at 150°C 3.4 ± K p.2 ↗ EDS056-230V
Temperature fluctuation at 150°C 0.8 ± K p.2 ↗ EDS056-230V
Heating up time to 150°C 60 min p.2 ↗ EDS056-230V
Recovery time after door was opened for 30 s at 150°C 45 min p.2 ↗ EDS056-230V
Rated voltage 230 V p.2 ↗ EDS056-230V
Power frequency 50/60 Hz p.2 ↗ EDS056-230V
Nominal power 1,05 kW p.2 ↗ EDS056-230V
Unit fuse 6,3 A p.2 ↗ EDS056-230V
Phase (nominal voltage) 1~ p.2 ↗ EDS056-230V
Interior volume 62 L p.2 ↗ EDS056-230V
Net weight of the unit (empty) 36 kg p.2 ↗ EDS056-230V
Load per rack 15 kg p.2 ↗ EDS056-230V
Permitted load 45 kg p.2 ↗ EDS056-230V
Wall clearance back 160 mm p.2 ↗ EDS056-230V
Wall clearance sidewise 100 mm p.2 ↗ EDS056-230V
Width net 614 mm p.2 ↗ EDS056-230V
Height net 625 mm p.2 ↗ EDS056-230V
Depth net 577 mm p.2 ↗ EDS056-230V
Interior width 400 mm p.2 ↗ EDS056-230V
Interior height 440 mm p.2 ↗ EDS056-230V
Interior depth 350 mm p.3 ↗ EDS056-230V
Unit doors 1 p.3 ↗ EDS056-230V
Energy consumption at 150°C 210 Wh/h p.3 ↗ EDS056-230V
Number of shelves (std./max.) 1/3 p.3 ↗ EDS056-230V

Operating Limits

  • Temperature range: +7 °C above ambient temperature to 250 °C p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗
  • Permitted load: 45 kg p.2 ↗
  • Load per rack: 15 kg p.2 ↗

Type Code Decoder

Position Token Meaning
1-3 p.1 ↗ EDS Product line: ED-S drying and heating chamber, Solid.Line
4-6 p.1 ↗ 056 Size / interior volume class (56, 62 L)
7-9 p.1 ↗ UL UL variant for 120 V / 60 Hz NEMA 5-15; absent for standard 230 V variant
10-13 p.1 ↗ 230V / 120V Rated voltage variant

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
EDS056-230V p.1 ↗ 9090-0014 230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard
EDS056UL-120V p.1 ↗ 9090-0015 120 V 1~ 60 Hz, NEMA 5-15 plug, Standard