Data Sheet Model MKF CO₂ 400

Technical Reference Summary

Covered Models & Document Scope

Cámara climática dinámica de CO₂ modelo MKF 400 con control de humedad, en dos variantes: MKF400CO2-400V (400 V 3~/N/PE 50 Hz, Art.-No. 9020-0519) y MKF400-480V (480 V 3~/N/PE 60 Hz, Art.-No. 9020-0521).

Technical Data

Parameter Value Condición
Temperature range -50...180 °C p.3 ↗ MKF400CO2-400V / MKF400-480V
Temperature uniformity dependent on set 0.1...1.0 ± K p.3 ↗ MKF400CO2-400V / MKF400-480V
Temperature fluctuation dependent on set 0.1...0.4 ± K p.3 ↗ MKF400CO2-400V / MKF400-480V
Cooling down time from 180°C to -40°C 80 min p.3 ↗ MKF400CO2-400V / MKF400-480V
Temperature range with humidity 10...95 °C p.3 ↗ MKF400CO2-400V / MKF400-480V
Humidity range 10...98 % RH p.3 ↗ MKF400CO2-400V / MKF400-480V
Max. heat compensation at 25°C and 90% RH 1,600 W p.3 ↗ MKF400CO2-400V / MKF400-480V
Rated voltage 400 V p.3 ↗ MKF400CO2-400V, 50 Hz
Rated voltage 480 V p.3 ↗ MKF400-480V, 60 Hz
Power frequency 50 Hz p.3 ↗ MKF400CO2-400V
Power frequency 60 Hz p.3 ↗ MKF400-480V
Nominal power 7,9 kW p.3 ↗ MKF400CO2-400V / MKF400-480V
Unit fuse 16,0 A p.3 ↗ MKF400CO2-400V / MKF400-480V
Phase (nominal voltage) 3~/N/PE p.3 ↗ MKF400CO2-400V / MKF400-480V
Interior volume 417 L p.3 ↗ MKF400CO2-400V / MKF400-480V
Net weight of the unit (empty) 394 kg p.3 ↗ MKF400CO2-400V
Net weight of the unit (empty) 404 kg p.3 ↗ MKF400-480V
Load per rack 30 kg p.3 ↗ MKF400CO2-400V / MKF400-480V
Permitted load 150 kg p.3 ↗ MKF400CO2-400V / MKF400-480V
Width net 1,115 mm p.4 ↗ MKF400CO2-400V / MKF400-480V
Height net 1,727 mm p.4 ↗ MKF400CO2-400V / MKF400-480V
Depth net 1,400 mm p.4 ↗ MKF400CO2-400V / MKF400-480V
Sound-pressure level 62 dB(A) p.4 ↗ MKF400CO2-400V / MKF400-480V
Energy consumption at 25°C and 60% RH 1,550 Wh/h p.4 ↗ MKF400CO2-400V / MKF400-480V
Number of shelves (std./max.) 1/6 p.4 ↗ MKF400CO2-400V / MKF400-480V

Operating Limits

  • Temperature range: -50...180 °C p.3 ↗
  • Temperature range with humidity: 10...95 °C p.3 ↗
  • Humidity range: 10...98 % RH p.3 ↗
  • Permitted load: 150 kg p.3 ↗
  • Load per rack: 30 kg p.3 ↗
  • Wall clearance back: 300 mm p.4 ↗

Type Code Decoder

Position Token Meaning
1-3 p.1 ↗ MKF Series: dynamic climate chamber with humidity control
4-6 p.1 ↗ 400 Model size (interior volume class 400)
suffix p.1 ↗ CO2 Refrigerant variant: natural CO₂ (R744)
voltage field p.2 ↗ 400V Mains variant 400 V 3~/N/PE 50 Hz (Art.-No. 9020-0519)
voltage field p.2 ↗ 480V Mains variant 480 V 3~/N/PE 60 Hz (Art.-No. 9020-0521)

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
MKF400CO2-400V p.2 ↗ 9020-0519 400 V 3~/N/PE 50 Hz, Standard version Plug: Standard
MKF400-480V p.2 ↗ 9020-0521 480 V 3~/N/PE 60 Hz, Standard version Plug: NEMA L22-20P

Modelos relacionados