Agilent Flow Cytometer Specifications
Technical Reference Summary
Covered Models & Document Scope
Especificaciones del citómetro de flujo Agilent NovoCyte que abarcan NovoCyte Penteon (5 láseres), NovoCyte Quanteon (4 láseres) y NovoCyte Advanteon (1–3 láseres)
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Laser type | Solid state laser with onboard thermal-electric cooling p.2 ↗ | |
| Laser beam profile | 10 × 60 μm p.2 ↗ | elliptical beam |
| Laser power | 405 nm 50 mW, 488 nm 60 mW, 561 nm 50 mW, 640 nm 40 mW mW p.2 ↗ | NovoCyte Advanteon |
| Fluorescence detection | Silicon photomultiplier (SiPM) with individual photodetector for each channel p.2 ↗ | |
| FSC sensitivity | 0.4 μm p.2 ↗ | |
| SSC sensitivity | 0.1 μm p.2 ↗ | |
| Fluorescence threshold sensitivity | FITC ≤40 MESF, PE ≤10 MESF, APC ≤10 MESF, Pacific Blue ≤30 MESF MESF p.2 ↗ | NovoCyte Quanteon/Penteon |
| Fluorescence resolution | <3 % CV p.2 ↗ | for CEN |
| Flow cell | 170 × 290 μm p.2 ↗ | rectangular quartz flow cell |
| Sample acquisition rate | 100,000 events/second p.2 ↗ | |
| Volumetric absolute count precision | <5 % p.2 ↗ | |
| Sample flow rate | 5 to 120 μL/min p.2 ↗ | continuously adjustable |
| Sheath flow rate | 6.5 mL/min p.2 ↗ | |
| Sample acquisition volume | 5 μL to 5 mL p.2 ↗ | |
| Manual sample format | 12 × 75 mm tube, 1.5 mL Eppendorf tube p.2 ↗ |