Centrifugation Brochure Centrifuge 5702 5702 R 5702 RH Its Your Turn.p
Technical Reference Summary
Covered Models & Document Scope
Familia Eppendorf Centrifuge 5702: Centrifuge 5702 (no refrigerada), Centrifuge 5702 R (refrigerada), Centrifuge 5702 RH (calefactada y refrigerada). También se cubren los rotores y portatubos (Rotor A-4-38, portatubos redondo, portatubos rectangular) y los accesorios.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Max. RCF | 3,000 × g p.11 ↗ | 5702, 5702 R, 5702 RH |
| Speed | 100–4,400 rpm p.11 ↗ | 5702, 5702 R, 5702 RH (100 rpm increments) |
| Max. capacity | 4 × 100 mL p.11 ↗ | 5702, 5702 R, 5702 RH |
| Rotor A-4-38 max. RCF | 3,052 × g p.9 ↗ | 4,400 rpm |
| Rotor A-4-38 max. RCF | 2,750 × g p.9 ↗ | 4,400 rpm |
| Rotor A-4-38 centrifugation radius | 134 mm p.8 ↗ | 1.5–2.0 mL round bottom |
| Rotor A-4-38 centrifugation radius | 132 mm p.8 ↗ | 1.1–1.4 mL, 2–7 mL, 2.6–7 mL, 4–10 mL |
| Rotor A-4-38 centrifugation radius | 136 mm p.8 ↗ | Eppendorf |
| Rotor A-4-38 centrifugation radius | 130 mm p.8 ↗ | 9–15 mL |
| Rotor A-4-38 centrifugation radius | 141 mm p.8 ↗ | 8 mL |
| Rotor A-4-38 centrifugation radius | 137 mm p.8 ↗ | 15 mL conical |
| Rotor A-4-38 centrifugation radius | 135 mm p.8 ↗ | 25 mL, 50 mL conical, 85–100 mL |
| Rotor A-4-38 centrifugation radius | 119 mm p.9 ↗ | 2–7 mL |
| Rotor A-4-38 centrifugation radius | 127 mm p.9 ↗ | 5–7 mL, 9 mL, 15 mL, 20 mL, 25 mL |
| Rotor A-4-38 max. RCF | 2,900 × g p.8 ↗ | 1.5–2.0 mL, 25 mL, 50 mL conical, 85–100 mL |
| Rotor A-4-38 max. RCF | 2,850 × g p.8 ↗ | 1.1–1.4 mL, 2–7 mL, 2.6–7 mL, 4–10 mL |
| Rotor A-4-38 max. RCF | 2,943 × g p.8 ↗ | Eppendorf |
| Rotor A-4-38 max. RCF | 2,800 × g p.8 ↗ | 9–15 mL |
| Rotor A-4-38 max. RCF | 3,052 × g p.8 ↗ | 8 mL |
| Rotor A-4-38 max. RCF | 3,000 × g p.8 ↗ | 15 mL conical |
| Rotor A-4-38 max. RCF | 2,577 × g p.9 ↗ | 2–7 mL |
| Rotor A-4-38 max. RCF | 2,750 × g p.9 ↗ | 5–7 mL, 9 mL, 15 mL, 20 mL, 25 mL |
| Acceleration time | 25 s p.11 ↗ | 5702, 5702 R, 5702 RH |
| Deceleration time | 25 s p.11 ↗ | 5702, 5702 R, 5702 RH |
| Noise level | <51 dB(A) p.11 ↗ | 5702 with Rotor A-4-38 |
Operating Limits
- Temperature settings (5702 R): –9 °C to 40 °C p.11 ↗
- Temperature settings (5702 RH): –9 °C to 42 °C p.11 ↗
- Timer: 0–99 min, with continuous run function p.11 ↗
- Max. power consumption (5702): 200 W p.11 ↗
- Max. power consumption (5702 R/RH): 380 W p.11 ↗
- Dimensions (W × D × H), 5702: 32 × 40 × 24 cm p.11 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | Aplica a | Note |
|---|---|---|---|---|
| 5702 745 006 p.8 ↗ | 5702 745 006 | Rotor A-4-38 round bucket, 1.5–2.0 mL, set of 2 | ||
| 5702 736 007 p.8 ↗ | 5702 736 007 | Rotor A-4-38 round bucket, 1.1–1.4 mL, set of 2 | ||
| 5702 737 003 p.8 ↗ | 5702 737 003 | Rotor A-4-38 round bucket, 2–7 mL, set of 2 | ||
| 5702 719 005 p.8 ↗ | 5702 719 005 | Rotor A-4-38 round bucket, 2.6–7 mL, set of 2 | ||
| 5702 735 000 p.8 ↗ | 5702 735 000 | Rotor A-4-38 round bucket, 4–10 mL, set of 2 | ||
| 5702 733 008 p.8 ↗ | 5702 733 008 | Rotor A-4-38 round bucket, Eppendorf, set of 2 | ||
| 5702 724 009 p.8 ↗ | 5702 724 009 | Rotor A-4-38 round bucket, 9–15 mL, set of 2 | ||
| 5702 764 000 p.8 ↗ | 5702 764 000 | Rotor A-4-38 round bucket, 8 mL, set of 2 | ||
| 5702 732 001 p.8 ↗ | 5702 732 001 | Rotor A-4-38 round bucket, 15 mL conical, set of 2 | ||
| 5702 723 002 p.8 ↗ | 5702 723 002 | Rotor A-4-38 round bucket, 15 mL conical, set of 2 |