Induztr
Resumen técnico Visor de documentos Modelos relacionados

Fermentors Bioreactors Brochure Parallel Bioreactor Systems Cell Cultu

Technical Reference Summary

Covered Models & Document Scope

Sistemas de biorreactores paralelos DASGIP y sistema de minibiorreactores DASbox: sistemas de 4, 8, 16 y 24 unidades, con o sin Bioblock, en variantes estándar (CC) y de un solo uso (CCSU); biorreactores de vidrio y de un solo uso; volúmenes de trabajo de 60–250 mL (vidrio) / 100–250 mL (un solo uso) para DASbox y de 250 mL–1.5 L (vidrio) / 320 mL–1.25 L (un solo uso) para DASGIP.

Technical Data

Parameter Value Condición
Number of parallel bioreactors up to 24 p.4 ↗ DASbox
Number of parallel bioreactors up to 16 p.4 ↗ DASGIP
Working volume 60 – 250 mL p.4 ↗ DASbox, glass
Working volume 100 – 250 mL p.4 ↗ DASbox, single-use
Working volume 250 – 1.5 L p.4 ↗ DASGIP, glass
Working volume 320 – 1.25 mL p.4 ↗ DASGIP, single-use
Drive Overhead drive p.4 ↗ DASbox and DASGIP
Impellers Marine (glass) / pitched blade (single-use) p.4 ↗ DASbox
Impellers Pitched blade p.4 ↗ DASGIP
Agitation speed range 20 – 2,500 rpm p.4 ↗ DASbox, glass
Agitation speed range 20 – 500 rpm p.4 ↗ DASbox, single-use
Agitation speed range 30 – 1,250 rpm p.4 ↗ DASGIP, standard
Agitation speed range 100 – 1,600 rpm p.4 ↗ DASGIP, alternative
Temperature control Liquid-free heating and cooling (Peltier) p.4 ↗ DASbox
Temperature control Integrated in the DASGIP Bioblock (optional) p.4 ↗ DASGIP
Standard temperature range 10 – 60 °C p.4 ↗ DASbox, at 25 °C RT
Standard temperature range 5 K above cooling agent temperature – 99 °C p.4 ↗ DASGIP
Feed lines per vessel 2 (standard) / 4 (optional) p.4 ↗ DASbox
Feed lines per vessel Up to 8 p.4 ↗ DASGIP
Standard feed rates 0.3 – 9.5 to 13 – 420 mL/h p.4 ↗ DASbox and DASGIP
Gas flow control TMFC; overlay and/or sparger p.4 ↗ DASbox and DASGIP
Standard gas mixing Air, N2, O2 and/or CO2 p.4 ↗ DASbox and DASGIP
Standard gas flow rates 0.04 – 5, 0.04 – 3.5 CO2 sL/h p.4 ↗ DASbox
Standard gas flow rates 0.1 – 50, 0.1 – 40 CO2 sL/h p.4 ↗ DASGIP
Software DASware control, other DASware optional p.4 ↗ DASbox and DASGIP

Operating Limits

  • Standard temperature range (DASbox): 10 – 60 °C at 25 °C RT p.4 ↗
  • Standard temperature range (DASGIP): 5 K above cooling agent temperature – 99 °C p.4 ↗
  • Maximum number of parallel bioreactors (DASbox): up to 24 p.4 ↗
  • Maximum number of parallel bioreactors (DASGIP): up to 16 p.4 ↗
  • Minimum agitation speed: 30 rpm p.3 ↗

Type Code Decoder

Position Token Meaning
1-2 p.4 ↗ 76 DASGIP/DASbox parallel bioreactor system family prefix
3-4 p.4 ↗ DX / DG DX = system without Bioblock; DG = system with Bioblock
5-6 p.4 ↗ 04 / 08 / 16 / 24 number of parallel units (4-, 8-, 16- or 24-fold system)
7-8 p.4 ↗ CC standard system configuration
9-10 p.4 ↗ SU single-use bioreactor system variant (suffix)
9-10 p.4 ↗ BB Bioblock variant (suffix, used with DG prefix)

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
76DX04CC p.4 ↗ 76DX04CC 4-fold system, standard
76DX04CCSU p.4 ↗ 76DX04CCSU 4-fold system, single-use
76DX08CC p.4 ↗ 76DX08CC 8-fold system, standard
76DX08CCSU p.4 ↗ 76DX08CCSU 8-fold system, single-use
76DX16CC p.4 ↗ 76DX16CC 16-fold system, standard
76DX16CCSU p.4 ↗ 76DX16CCSU 16-fold system, single-use
76DX24CC p.4 ↗ 76DX24CC 24-fold system, standard
76DX24CCSU p.4 ↗ 76DX24CCSU 24-fold system, single-use
76DG04CCBB p.4 ↗ 76DG04CCBB 4-fold system with Bioblock
76DG04CCSU p.4 ↗ 76DG04CCSU 4-fold system with Bioblock, single-use