CC-200BX Cooling Thermostat Datasheet

Contenido

  1. Overview p. 1
    Product introduction and main features of the CC-200BX cooling thermostat
  2. Spec p. 2
    Technical data according to DIN 12876, including temperature and performance figures
  3. p. 1 table, 35 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range28...200 °C
with water cooling20...200 °C
with refrigerator-20...200 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
Pressure pumpyes
max. delivery27 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)22 l/min
max. delivery pressure (suction)0,4 bar
Pump connenction (optional)M16x1 male
Immersion depth128 mm
max. permissible kin. viscosity50 mm²/s
Overall dimensions WxDxH **345x200x326 mm
Net weight7 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
alternative max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method: 28–200 °C without cooling, 20–200 °C with water cooling, and -20–200 °C with a refrigerator.

Determines whether the unit can reach the required process temperature with the available cooling utility.

Source
“Operating temperature range 28...200 °C with water cooling 20...200 °C with refrigerator -20...200 °C”
Installation

The unit requires a 200–240 V single- or two-phase supply at 50/60 Hz with a maximum current of 10 A and a fuse rated between 10 A and 16 A.

Needed to verify site power availability and correct branch-circuit protection before installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A alternative max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation

Voltage and frequency tolerances are ±10% only if they do not run in opposite directions, so a -10% voltage with +3% frequency is not allowed.

Prevents installation on a supply where combined voltage and frequency deviations exceed the permitted envelope.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Safety

External branch circuit protection according to UL 489 is required.

Specifies the type of upstream protection that must be provided for compliant installation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.

Ensures the correct plug is ordered for the destination country and avoids rework at commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

The ambient temperature must stay between 5 °C and 40 °C, and the unit has IP20 protection, so it must be kept in a dry indoor location.

Defines the permissible site environment and rules out wet or dusty locations.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

The pressure pump delivers up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar, which limits the connected external circuit.

Used to check that the external loop pressure drop and flow demand are within pump capability.

Source
“max. delivery 27 l/min max. delivery pressure 0,7 bar Suction pump yes max. delivery (suction) 22 l/min max. delivery pressure (suction) 0,4 bar”
Performance

The maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s.

Restricts the choice of heat transfer fluid, especially at low temperatures where viscosity rises.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The immersion depth is 128 mm and the bridge thermostat has a telescopic arm up to 884 mm, which sets the bath mounting geometry.

Needed to confirm the bath depth and mounting position allow proper immersion of the heater and sensor.

Source
“Immersion depth 128 mm”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.

Confirms the safety concept required for flammable or sensitive bath fluids and for unattended operation.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Cumplimiento

The unit is classified as Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment may require additional EMC measures in residential or light-industrial environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Selection

The pressure equipment category is Art. 4.3 PED, which affects the documentation and conformity route for the pressure side.

Relevant for compliance assessment of the pressure-containing circuit under the PED.

Source
“Pressure equipment category Art. 4.3 PED”
Operation

The temperature stability at 70 °C is 0.02 K, which sets the achievable bath uniformity for sensitive processes.

Used to judge whether the unit meets the process tolerance for temperature-sensitive applications.

Source
“Temperature stability at 70°C 0,02 K”
Maintenance

The housing and all moistened parts are stainless steel, and the pressure and suction pumps are made of high-resistant plastic.

Informs fluid compatibility and cleaning/maintenance practice for wetted components.

Source