Technical data according to DIN 12876

ParameterValue
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:402842

Technical data according to DIN 12876

ParameterValue
Operating temperature range45...200 °C
with water cooling20...200 °C
with refrigerator-30...200 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
Pressure pumpyes
max. delivery27 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)22 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
Bath volume3,5 l
Filling capacity2 l
Bath opening diameter25 mm
Bath depth150 mm
Height of bath opening190 mm
Overall dimensions WxDxH **178x260x355 mm
Net weight9 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
alternative max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method: 45–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator.

Determines whether the unit can reach the required process temperature with the available cooling utility.

Source
“Operating temperature range 45...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Installation

The unit requires a 200–240 V single- or two-phase supply with a minimum 10 A fuse and maximum 16 A fuse.

Electrical supply and branch protection must be sized accordingly before installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A alternative max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation

External branch circuit protection according to UL 489 is required.

The installer must provide a UL 489-listed branch breaker; this affects panel and wiring design.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are ±10% only if they do not run in opposite directions; for example, -10% voltage with +3% frequency is not allowed.

Site power quality must be checked; combined opposite tolerances can make the supply non-compliant.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.

The correct plug type must be selected on the order to avoid rework at commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation

The unit needs a water cooling coil connection of 3/8 inch for tap water cooling.

Cooling water piping and fittings must be prepared for a 3/8" connection if water cooling is used.

Source
“Cooling coil for (tap) water (3/8").”
Installation

The pump connection is M16x1 male, and sleeve nuts with M16x1 thread are standard equipment.

External circuit hoses and adapters must match the M16x1 male thread.

Source
“Pump connection M16x1 male”
Operation

The permissible ambient temperature range is 5 °C to 40 °C.

The installation location must stay within this range for reliable operation and performance.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

The pressure pump delivers up to 27 l/min at 0.7 bar, while the suction pump delivers up to 22 l/min at 0.4 bar.

External circuit pressure drop and required flow must be checked against these pump limits.

Source
“max. delivery 27 l/min max. delivery pressure 0,7 bar Suction pump yes max. delivery (suction) 22 l/min max. delivery pressure (suction) 0,4 bar”
Selection

The bath volume is 3.5 l with a filling capacity of 2 l, and the bath opening is 25 mm diameter by 150 mm deep.

Determines whether the bath opening and volume suit the immersion depth and sample size.

Source
“Bath volume 3,5 l Filling capacity 2 l Bath opening diameter 25 mm Bath depth 150 mm”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.

Overtemperature protection must be set correctly for the application to meet safety requirements.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Cumplimiento

The unit is classified as Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment may require additional EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The degree of protection is IP20, so the unit must be protected from water ingress and dust in the installation environment.

IP20 limits where the unit can be installed and requires protection from splashes and dust.

Source
“Degree of Protection IP20”
Selection

The pressure equipment category is Art. 4.3 PED.

Affects the pressure equipment compliance documentation and assessment required for the installation.

Source
“Pressure equipment category Art. 4.3 PED”