Selection
The unit's operating temperature range depends on the cooling method chosen, so the cooling option must be specified at selection.
Determines whether the bath can reach the required process temperatures without an external cooler.
Source
“Operating temperature range 25...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Installation
The electrical supply must be 200-240 V single or two-phase at 50/60 Hz with a branch circuit rated between 10 A and 16 A.
Needed to size the site power feed and protective device correctly before installation.
Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation
External branch circuit protection to UL 489 is mandatory for this equipment.
The installer must provide a listed UL 489 breaker; the unit does not include branch protection.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are each ±10% but must not deviate in opposite directions simultaneously.
Prevents connecting the unit to a supply where combined voltage/frequency drift exceeds the allowed envelope.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
Single- and two-phase units are delivered with a power cable and country-specific plug that must be specified on the order.
Avoids receiving a unit with the wrong plug for the site and delays during commissioning.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The bath requires a minimum filling of 5 l within its 7.5 l volume.
Underfilling risks pump cavitation and poor temperature control; overfilling risks overflow at temperature.
Source
“Bath volume 7,5 l min. filling capacity 5 l”
Operation
The pressure pump delivers up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar.
Used to check whether the external loop's flow and pressure drop are within the pump capability.
Source
“max. delivery 27 l/min max. delivery pressure 0,7 bar”
Operation
The bath bridge has a dedicated hole for a cooling probe, allowing an immersion cooler to be fitted.
Enables extending the low-temperature range by adding an immersion cooler without modifying the bath.
Source
“Bath bridge with hole for cooling probe (e.g. for immersion cooler TC45-TC100E).”
Operation
When a Huber pump adapter is used, polyglycol must not be used as the heat transfer fluid.
Prevents selecting an incompatible fluid that could damage the pump adapter or compromise safety.
Source
“When using Huber pump adapter: Polyglycol is not permissible to be used as a heat transfer fluid”
Safety
The unit has adjustable overtemperature protection per DIN 12876 and safety classification III / FL.
Confirms the safety concept required for flammable or sensitive fluids and for unattended operation.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Cumplimiento
The equipment is classified as pressure equipment under Article 4.3 PED.
Affects documentation and conformity obligations when placing the unit into service in the EU.
Source
“Pressure equipment category Art. 4.3 PED”
Cumplimiento
The unit is EMC Class A, Group 1 to EN55011, so it is intended for non-domestic environments.
Class A equipment may cause interference in residential areas; site EMC planning must account for this.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient.
Defines the required room conditions and excludes wet or dusty locations from installation.
Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Maintenance
A drain valve with cap is available as an accessory for emptying the bath.
Needed for safe fluid changes and maintenance; should be ordered with the unit if draining is required.
Source
“Drain valve with cap #6839”