CC-K25 Refrigerated/Heating Circulator

Contenido

  1. Overview p. 1
    Product name and general description of the CC-K25 unit
  2. Spec pp. 1-2
    Technical data and performance parameters per DIN 12876
  3. p. 2 table, 12 rows
  4. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
Net weight41 kg
sound pressure level +/- 4 dB(A)50 dB(A)
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current refrigerated bath2 A
max. current immersion thermostat10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S395642

Technical data according to DIN 12876

ParameterValue
Operating temperature range-30...200 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at 20°C0,4 kW
Cooling power at 0°C0,35 kW
Cooling power at -10°C0,27 kW
Cooling power at -20°C0,16 kW
Cooling power at -30°C0,065 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,072 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Pressure / Suction pump max. delivery27 l/min
max. delivery pressure0,7 bar
max. delivery (suction)22 l/min
max. delivery pressure (suction)0,4 bar
Pump connenction (optional)M16x1 male
Bath volume23,5 l
Width bath opening WxD290x329 mm
Bath depth200 mm
Height of bath opening450 mm
Overall dimensions WxDxH **350x555x615 mm

** см. сноску в оригинале (на странице не приведена)

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is classified as safety class III / FL per DIN 12876, which determines the required overtemperature protection and suitability for flammable liquids.

Safety class and FL designation dictate whether the bath can be used with flammable heat transfer fluids and what protective measures are mandatory.

Source
“Safety classification III / FL”
Selection

The refrigeration circuit uses R-290 (propane, A3, H220) with a charge of only 0.072 kg, so the installation location must satisfy flammable-refrigerant requirements.

A3 refrigerant affects room ventilation, ignition-source restrictions and service access during installation and maintenance.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Installation

The power supply must be 220-240 V single- or two-phase at 50/60 Hz, and the branch circuit must be protected with a fuse between 10 A and 16 A.

Defines the electrical supply and upstream protection needed before the unit can be connected.

Source
“Power supply requirement 220-240V 1~/2~ 50/60Hz”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is explicitly not allowed.

Prevents commissioning on a supply where the combined deviation would exceed the motor's permissible operating envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The enclosure is rated IP20, so the unit must be installed in a dry location protected from water ingress.

IP20 rules out washdown or wet environments and constrains where the bath can be sited.

Source
“Degree of Protection IP20”
Installation

Ambient temperature at the installation site must stay between 5 °C and 40 °C.

Outside this range cooling capacity drops and reliable operation is not guaranteed.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity falls from 0.4 kW at 20 °C to 0.065 kW at -30 °C, so the achievable heat load depends strongly on the working temperature.

Sizing the bath for a process requires checking the cooling capacity at the actual setpoint, not at 20 °C.

Source
“at 20°C 0,4 kW at 0°C 0,35 kW at -10°C 0,27 kW at -20°C 0,16 kW at -30°C 0,065 kW”
Performance

Cooling capacity is specified at 20 °C room temperature and may drop if ambient temperature rises.

Prevents overestimating available cooling in a warm laboratory and guides ventilation planning.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation

The pressure/suction pump delivers up to 27 l/min at 0.7 bar pressure and 22 l/min at 0.4 bar suction.

Determines whether the pump can supply an external circuit or cascade at the required flow and head.

Source
“max. delivery 27 l/min max. delivery pressure 0,7 bar max. delivery (suction) 22 l/min max. delivery pressure (suction) 0,4 bar”
Operation

When a Huber pump adapter is used, polyglycol must not be used as the heat transfer fluid.

Prevents selecting an incompatible fluid that would damage the pump adapter or compromise safety.

Source
“When using Huber pump adapter: Polyglycol is not permissible to be used as a heat transfer fluid”
Cumplimiento

The equipment is classified as Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment is intended for industrial environments and may require EMC mitigation if installed near sensitive laboratory instruments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at order.

The correct plug type must be stated on the order, otherwise the unit cannot be connected on arrival.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Maintenance

The unit is fitted with a drain valve with cap as an accessory, which is needed for safe emptying of the 23.5 l bath.

Planning fluid changes and disposal requires the drain valve to be ordered and accessible.

Source
“Drain valve with cap #6839”
Safety

The bath has adjustable overtemperature protection according to DIN 12876.

The overtemperature cut-out must be set correctly for the fluid used to prevent overheating and fire risk.

Source
“With adjustable overtemperature protection according to DIN 12876.”