Product identification and general description of the CS 330 OD unit
Specp. 2
Technical data table according to DIN 12876 standard
p. 1table, 40 rows
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-20...15 °C
temperature set point / display
colour LED Touchscreen
Internal temperature sensor
Pt100
Temperature stability at -10°C
1,5 K
Interface digital
RS232
Safety classification
I / NFL
Kälteleistung bei Umgebungstemperatur 5°C
at 15°C
40 kW
at 10°C
32 kW
at 0°C
22 kW
at -10°C
15 kW
at -20°C
10 kW
Cooling power at ambient temperature 20°C
at 15°C
42 kW
at 10°C
35 kW
at 0°C
25 kW
at -10°C
17 kW
at -20°C
11 kW
Kälteleistung bei Umgebungstemperatur 35°C
at 15°C
40 kW
at 10°C
33 kW
at 0°C
22 kW
at -10°C
15 kW
at -20°C
9,5 kW
Kälteleistung bei Umgebungstemperatur 40°C
at 15°C
38 kW
at 10°C
27 kW
at 0°C
21 kW
at -10°C
13 kW
at -20°C
7,5 kW
Kälteleistung bei Umgebungstemperatur 50°C
at 15°C
31 kW
at 10°C
26 kW
at 0°C
18 kW
at -10°C
11 kW
at -20°C
6,5 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Refrigerant quantity
9 kg
CO2 equivalent
12,57 t
Engineering Facts
12 facts · every claim linked to a page
Selection
Cooling capacity must be selected based on the actual ambient temperature at the installation site, since capacity drops significantly at higher ambient temperatures.
Prevents undersizing the chiller when installed in a warm environment.
Source
“Output data valid for: Room temperature 20°C to 50°C . If the ambient temperature rises, the cooling capacity may drop.”
Installation
The chiller is suitable for outdoor installation down to -20°C ambient, but cooling capacity is reduced above +35°C ambient.
Determines whether the unit can be placed outdoors and what capacity derating to expect.
Source
“Suitable for outside mounting (min. environmental temperature -20°C)”
Installation
Three-phase devices under 63A are supplied with cable but without a plug, so the installer must provide a suitable plug or hardwired connection.
Affects electrical installation planning and bill of materials.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances are interdependent: -5% voltage with +2% frequency is not allowed, but -5% voltage with -2% frequency is allowed.
Ensures the site power quality is compatible with the chiller's EN60034-1 tolerances.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation requirements apply.
Needed for refrigerant safety assessment and compliance with local regulations.
Source
“R-449A (A1, H280)”
Cumplimiento
The equipment emits Class A, Group 1 disturbances per EN55011, which may require EMC precautions in residential or sensitive environments.
Determines EMC compatibility and possible filtering or separation requirements on site.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation
The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.
Allows matching pump flow to the process without external throttling.
Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Operation
The remote control panel is connected via a 20 m cable, allowing the operator interface to be placed away from the chiller.
Affects control room layout and cable routing during installation.
Source
“Delivered with remote control: operating panel connected via 20m cable”
Maintenance
The recommended thermofluid is a 50:50 water-monoethylene glycol mixture.
Ensures correct fluid selection for freeze protection and heat transfer performance.
Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Installation
Space requirements around the unit must be respected according to the operating conditions referenced on the manufacturer's website.
Prevents airflow restriction and ensures adequate clearance for air-cooled condenser.
Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Selection
The chiller uses an air-cooled refrigeration machine with CFC- and HCFC-free refrigerant.
Confirms refrigerant compliance and that no water cooling circuit is required.
Source
“air-cooled, CFC- and HCFC-free”
Safety
The electrical components have protection classification IP54, suitable for dusty and splash-water environments.
Determines suitability for outdoor or harsh installation locations.
Source
“Protection classification of the electrical components IP54”