Contenido

  1. Overview p. 1
    Product introduction and main features of KISS 106A
  2. Spec p. 2
    Specifications according to DIN 12876 standard
  3. p. 1 table, 36 rows

Technical data according to DIN 12876 — KISS 106A

ParameterValue
Operating temperature range25...100 °C
with water cooling20...100 °C
with refrigerator15...100 °C
Temperature stability at 70°C0,05 K
temperature set point / displaydigital
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Interface digitalUSB (Device), RS232
InterfaceAlarm message
optic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
max. delivery14 l/min
max. delivery pressure0,25 bar
max. delivery (suction)10,5 l/min
max. delivery pressure (suction)0,17 bar
Pump connenction (optional)M16x1 male
Bath volume4,4 l
min. filling capacity2,5 l
Height of bath opening160 mm
Width bath opening WxD/ bath depth130x110/ 150 mm
Overall dimensions WxDxH **147x307x330 mm
Net weight5 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryN/A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.407137

** Overall dimensions WxDxH

Engineering Facts

13 facts · every claim linked to a page
Selection

The usable temperature range depends on the cooling method, so the chiller/cooling option must be chosen to reach the required low temperature.

Determines whether the bath can reach the process temperature without an external refrigerator or water cooling.

Source
“Operating temperature range 25...100 °C with water cooling 20...100 °C with refrigerator 15...100 °C”
Performance

Heating power is voltage-dependent, so the actual available heating capacity must be checked against the site supply voltage.

Affects heat-up time and whether the unit meets the process requirement at the installed voltage.

Source
“Heating power at 230V 2 kW”
Installation

The unit requires a 200-240 V single- or two-phase supply with a 10 A minimum and 16 A maximum fuse.

Needed to size the branch circuit and protective device correctly.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation

An external branch circuit protection according to UL 489 is mandatory.

Must be provided in the installation for compliance and safe operation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are limited and must not deviate in opposite directions.

Prevents specifying the unit on a supply where combined voltage/frequency deviation is out of tolerance.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Safety

The device is safety class III / FL with adjustable overtemperature protection per DIN 12876.

Relevant for risk assessment and for setting the overtemperature cut-out during commissioning.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The autostart function restarts the unit automatically after a power failure.

Important for unattended processes and for deciding whether this behavior is acceptable.

Source
“Autostart function for power failure”
Selection

The plastic pump is not resistant to acetone and polyglycols, so these media require an alternative pump material.

Prevents selecting the standard unit for incompatible bath fluids.

Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.

Guides the choice of heat transfer fluid for low-temperature operation.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The Pt100 process sensor option is factory-fitted only and is for display, not control.

Must be ordered with the unit and cannot be used for closed-loop control of process temperature.

Source
“Option: Pt100 sensor connection #10688 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge).”
Installation

The unit is IP20 and must be installed in an environment between 5 °C and 40 °C.

Defines the permissible installation location and ambient conditions.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

Overall dimensions and space requirements must be respected at the installation site.

Ensures adequate clearance and access for installation and operation.

Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Selection

The bath holds 4.4 l with a minimum filling capacity of 2.5 l.

Determines the required fluid volume and whether the bath suits the application.

Source
“Bath volume 4,4 l min. filling capacity 2,5 l”