Product introduction and key features of KISS 215B
Specp. 2
Technical data according to DIN 12876
p. 1table, 36 rows
Technical data according to DIN 12876
Technical data
KISS 215B
Operating temperature range
25...200 °C
with water cooling
20...200 °C
with refrigerator
-30...200 °C
Temperature stability at 70°C
0,05 K
temperature set point / display
digital
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
Interface digital
USB (Device), RS232
Interface
Alarm message
optic, acoustic
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
max. delivery
14 l/min
max. delivery pressure
0,25 bar
max. delivery (suction)
10,5 l/min
max. delivery pressure (suction)
0,17 bar
Pump connenction (optional)
M16x1 male
Bath volume
15 l
min. filling capacity
10 l
Height of bath opening
255 mm
Width bath opening WxD/ bath depth
290x152/ 200 mm
Overall dimensions WxDxH **
350x375x425 mm
Net weight
12 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
N/A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
398444
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The usable temperature range depends on the cooling method: 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator.
Determines whether the bath can reach sub-ambient or near-ambient setpoints with the available cooling utility.
Source
“Operating temperature range 25...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Selection
The bath holds 15 l but requires a minimum filling of 10 l, so the working volume and immersion depth must be planned accordingly.
Affects sample/immersion cooler fit and whether the bath can be operated safely at the intended fill level.
Source
“Bath volume 15 l min. filling capacity 10 l”
Installation
The unit requires a 200–240 V single- or two-phase 50/60 Hz supply with a minimum 10 A fuse and maximum 16 A fuse.
Sizing the branch circuit, plug, and protective device must match these limits before installation.
Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation
An external branch circuit protection device according to UL 489 is required.
The installer must provide a listed branch-circuit breaker; the unit does not include this protection.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are each ±10%, but they must not run in opposite directions (e.g., -10% voltage with +3% frequency is not allowed).
Prevents selecting a supply or generator whose combined voltage/frequency deviation exceeds the allowed envelope.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant plastic must be ordered.
Media compatibility determines whether the standard pump can be used or a special version is needed.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Prevents freezing and damage when operating near or below 0 °C with water-based fluid.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Safety
The unit is safety class III / FL with adjustable overtemperature protection according to DIN 12876.
Confirms the safety architecture and that overtemperature protection must be set for the application.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Safety
The autostart function restarts the unit after a power failure, which may be undesirable for some processes.
Engineers must decide whether automatic restart after mains loss is acceptable for their process.
Source
“Autostart function for power failure”
Installation
The bath bridge has a hole for a cooling probe, e.g., for an immersion cooler TC45–TC100E.
Allows planning of an immersion cooler for sub-ambient operation without modifying the bath bridge.
Source
“Bath bridge with hole for cooling probe (e.g. for immersion cooler TC45-TC100E).”
Installation
The Pt100 sensor connection option is factory-fitted only and displays (not controls) process temperature.
Must be ordered with the unit; it cannot be retrofitted in the field and does not provide closed-loop control.
Source
“Option: Pt100 sensor connection #10688 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge).”
Cumplimiento
The equipment is classified for electromagnetic disturbance as EN 55011 Class A, Group 1.
Class A equipment may require additional EMC measures in residential or light-industrial environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at order.
The correct plug type must be stated on the order to avoid a mismatch with the site receptacle.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The ambient temperature must remain between 5 °C and 40 °C during operation.
Site conditions outside this range can affect performance, safety, and warranty.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”