Contenido

  1. Overview p. 1
    Product introduction and key features of KISS E circulator
  2. Spec p. 2
    Technical specifications according to DIN 12876 standard
  3. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterKISS E
Operating temperature range25...200 °C
with water cooling20...200 °C
with refrigerator-30...200 °C
Temperature stability at 70°C0,05 K
temperature set point / displaydigital
Internal temperature sensorPt100
Interface digitalUSB (Device), RS232
Interface
Alarm messageoptic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
max. delivery14 l/min
max. delivery pressure0,25 bar
max. delivery (suction)10,5 l/min
max. delivery pressure (suction)0,17 bar
Pump connenction (optional)M16x1 male
Immersion depth150 mm
Overall dimensions WxDxH **132x163x312 mm
Net weight4 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
Power input2,1 kW
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryN/A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.S384165

** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Engineering Facts

14 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method, so the required range must be matched to whether water cooling or a refrigerator is used.

Determines whether the unit can cover the process temperature window without additional cooling equipment.

Source
“Operating temperature range 25...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Installation

The unit requires a 200-240 V single- or two-phase supply at 50/60 Hz and draws up to 10 A, so the site must provide a suitable circuit.

Needed to confirm site power availability and correct branch circuit design before installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz”
Installation

The minimum fuse must be 10 A and the maximum fuse 16 A, which constrains the upstream protection device selection.

Prevents nuisance tripping or inadequate protection of the supply circuit.

Source
“min. Fuse 10A max. Fuse 16A”
Installation

External branch circuit protection according to UL 489 is required, so the installer must provide a listed branch breaker.

Affects electrical installation compliance and hardware selection for the branch circuit.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are ±10% only if they do not run in opposite directions, so supply quality must be checked before commissioning.

Avoids operating the unit outside its allowed supply envelope and prevents damage or malfunction.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant pump specified.

Prevents chemical attack and premature pump failure when selecting the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Guides correct fluid preparation to avoid freezing and maintain safe operation.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Safety

The unit has adjustable overtemperature protection and safety class III/FL, which is relevant for unattended operation and risk assessment.

Supports safety classification and configuration of over-temperature limits for the process.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

An autostart function for power failure is included, so the behavior after a power interruption must be considered for the process.

Affects restart strategy and risk of unexpected heating or pumping after power returns.

Source
“Autostart function for power failure”
Installation

The degree of protection is IP20, so the unit must be installed in a dry, protected environment.

Restricts installation location and protects against dust and water ingress.

Source
“Degree of Protection IP20”
Installation

The ambient temperature must remain between 5 °C and 40 °C for proper operation.

Ensures the unit is not installed in an environment outside its rated operating conditions.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

The pump delivers up to 14 l/min with a maximum delivery pressure of 0.25 bar, which limits the external circuit resistance.

Used to check whether the pump can overcome the pressure drop of the external loop.

Source
“max. delivery 14 l/min max. delivery pressure 0,25 bar”
Installation

The immersion depth is 150 mm, so the thermostat must be mounted so that this depth is respected in the bath.

Determines mechanical mounting and ensures correct temperature sensing and pump operation.

Source
“Immersion depth 150 mm”
Cumplimiento

The disturbance classification is EN55011 Class A Group 1, which has implications for electromagnetic compatibility in the installation environment.

Class A equipment may require additional EMC measures depending on the installation location.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”