Induztr
Resumen técnico Visor de documentos Modelos relacionados

KISS K20 Immersion Circulator

Contenido

  1. Overview p. 1
    Product cover and model identification for KISS K20
  2. Spec p. 2
    Technical data table according to DIN 12876
  3. p. 2 table, 7 rows

Technical data according to DIN 12876

ParameterValue
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S393794

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit covers a wide operating temperature range from -30 to 200 °C, making it suitable for both sub-ambient and high-temperature bath applications.

Defines the process window the engineer can cover with a single unit, avoiding a separate chiller for low temperatures.

Source
“Operating temperature range -30...200 °C”
Performance

Cooling capacity drops sharply at low temperatures, from 0.4 kW at 20 °C to only 0.065 kW at -30 °C, so heat loads must be sized against the actual operating temperature.

Prevents oversizing assumptions based on nominal cooling power and ensures the bath can hold setpoint under load at the coldest point.

Source
“at -30°C 0,065 kW”
Safety

The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, which affects room ventilation, ignition-source control and refrigerant handling procedures.

Flammable refrigerant classification drives installation location, ventilation and safety documentation requirements.

Source
“R-290 (A3, H220)”
Installation

The refrigerating unit is air-cooled, so adequate air flow and clearance around the unit are required; space requirements must be checked before siting.

Air-cooled condensers lose capacity and can trip on high pressure if installed in tight or poorly ventilated spaces.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Installation

The unit requires a 220-240 V single- or two-phase 50/60 Hz supply and draws up to 10 A, so the circuit and plug must be rated accordingly.

Determines the electrical connection, breaker sizing and whether a country-specific plug must be specified at order.

Source
“Power supply requirement 220-240V 1~/2~ 50/60Hz”
Installation

Voltage and frequency tolerances are limited to ±5 % and ±2 % respectively, and the combination of -5 % voltage with +2 % frequency is explicitly not allowed.

Prevents installation on unstable or generator-fed supplies where simultaneous voltage and frequency deviations could damage the compressor.

Source
“-5% voltage and +2% frequency -> not allowed!”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant plastic pump ordered.

Prevents chemical attack and pump failure when selecting the heat transfer fluid for the application.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection the manufacturer recommends water mixed with glysantine or ethylene glycol rather than pure water.

Guides fluid selection to avoid freezing and damage when operating below ambient or near 0 °C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The unit is rated IP20, so it must be installed in a dry, protected location and not exposed to water spray or condensation.

IP20 limits the environment of installation and rules out washdown or wet locations.

Source
“Degree of Protection IP20”
Installation

Ambient temperature must stay between 5 and 40 °C, and cooling capacity may drop if ambient rises above 20 °C.

Room conditions directly affect achievable cooling and must be considered when sizing and siting the unit.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Cumplimiento

The equipment is EMC-classified as Class A, Group 1 to EN55011, which is intended for industrial environments and may cause interference in residential areas.

Class A equipment has implications for installation in non-industrial or residential environments and for EMC coordination.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The unit provides adjustable overtemperature protection according to DIN 12876 and is classified Safety class III / FL.

Confirms the safety concept required for unattended operation and for protection of temperature-sensitive samples.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

An autostart function restarts the unit after a power failure, which may be undesirable for some processes and should be evaluated.

Autostart behavior after mains interruption must be considered for process safety and unattended operation.

Source
“Autostart function for power failure”
Selection

A Pt100 sensor connection is available only factory-fitted as an option and is used to display, not control, an external process temperature.

Clarifies that external temperature monitoring is display-only and must be ordered with the unit, not retrofitted.

Source
“Option: Pt100 sensor connection #10688 to display (not control)”