Selection
The unit uses flammable refrigerant R-1270 (A3, H220), so the installation must respect flammable-refrigerant safety requirements.
A3 flammability affects room classification, ventilation, ignition-source control and local approval during selection and installation.
Source
“Refrigerant (ASHRAE, GHS) R-1270 (A3, H220)”
Installation
The power supply must be 208-240 V single/two-phase 50/60 Hz with a maximum current of 12.5 A and a 16 A fuse.
Defines the electrical feed, breaker/fuse rating and cabling needed before the chiller is installed.
Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but they must not be combined in the unfavorable direction.
Prevents selecting a supply point whose combined voltage/frequency deviation would exceed the allowed operating envelope.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
The unit is IP20, so it must be installed in a dry indoor location protected from water and dust ingress.
IP20 rules out washdown or outdoor installation and dictates enclosure/room protection.
Source
“Degree of Protection IP20”
Operation
Ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may drop above 20 °C.
Room conditions directly limit available cooling capacity and must be checked against the process heat load.
Source
“max. ambient temperature 40 °C”
Performance
Cooling capacity falls from 1 kW at +15 °C to 0.35 kW at -20 °C, so the required setpoint determines the usable capacity.
Sizing must be based on capacity at the actual operating temperature, not the headline figure.
Source
“at -20°C 0,35 kW”
Selection
The plastic pump is not resistant to acetone or polyglycols, so these fluids require a different pump material.
Fluid compatibility must be confirmed before ordering; otherwise a more resistant plastic option is needed.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.
Prevents freezing and damage when operating near or below 0 °C.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.
Affects electrical installation planning and the parts the customer must supply.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Cumplimiento
The equipment is EMC Class A Group 1, which is intended for industrial environments and may cause interference in residential areas.
Class A limits where the unit may be installed without additional EMC measures.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety
The unit has adjustable overtemperature protection per DIN 12876 and safety classification III/FL.
The safety cut-out must be set correctly for the bath fluid and process to protect against overheating.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Maintenance
The refrigerant charge is only 0.07 kg with GWP 0, so refrigerant handling and leak reporting obligations are minimal.
Reduces regulatory burden and service logistics for the refrigeration circuit.
Source
“Refrigerant quantity 0,07 kg”