Minichiller 1200-H OLÉ Technical Data Sheet

Contenido

  1. Overview p. 1
    Product name and cover page of the datasheet
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 2 table, 9 rows

Technical data according to DIN 12876

CharacteristicValue
Net weight38 kg
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current12 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S616906

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is classified as safety class III / FL per DIN 12876, which determines the required safety measures and permissible fluids for the application.

Safety class governs installation requirements and fluid selection for the temperature control application.

Source
“Safety classification III / FL”
Selection

The refrigerant is R-1270 (A3, H220) with GWP 0, so the unit uses a flammable natural refrigerant requiring appropriate site precautions.

Flammable A3 refrigerant affects room ventilation, ignition source control and risk assessment at the installation site.

Source
“R-1270 (A3, H220)”
Performance

Cooling capacity drops from 1.2 kW at 15 °C to 0.35 kW at -20 °C, so the unit must be sized against the actual operating temperature.

Selecting on nominal capacity alone will undersize the chiller at low set points.

Source
“at -20°C 0,35 kW”
Installation

The pump connections are M16x1 male, so mating hoses and fittings must match this thread.

Thread mismatch prevents connection; included sleeve nuts are M16x1.

Source
“Pump connection M16x1 male”
Operation

Whisper mode reduces sound level to 51 dB(A) but also cuts pump delivery to 14 l/min and 0.2 bar, so flow requirements must be checked before enabling it.

Reduced pump performance in whisper mode may be insufficient for the process heat load.

Source
“Delivery rate: 14 l/min”
Installation

The electrical supply must be 208-240 V single or two-phase 50/60 Hz with a 16 A fuse and max current 12 A.

Site power and protection must be provisioned accordingly before commissioning.

Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

Supply quality at the site must be verified to avoid invalid operating conditions.

Source
“-5% voltage and +2% frequency -> not allowed!”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant pump ordered.

Fluid compatibility determines pump life and prevents leaks or failure.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Correct fluid mixture prevents freezing and damage in the cooling circuit.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The unit has IP20 protection and must be operated between 5 °C and 40 °C ambient, so it cannot be placed in wet or hot environments.

IP20 and ambient limits restrict installation location and enclosure requirements.

Source
“Degree of Protection IP20”
Performance

Cooling capacity data is valid at 20 °C room temperature and may drop if ambient temperature rises.

Hot rooms will reduce available cooling and may require a larger unit.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

Plug type must be stated on the order to match site receptacles.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Cumplimiento

The equipment is EMC class A group 1 per EN55011, so it may cause radio interference in residential areas.

Class A equipment is intended for industrial environments, not residential use.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

A drain valve is an optional accessory, so gravity drainage for fluid changes requires ordering it separately.

Without the drain valve, fluid replacement and maintenance are more difficult.

Source
“Optional accessories: Drain valve #6839”