Induztr
Resumen técnico Visor de documentos Modelos relacionados

Minichiller 1200w-H Datasheet

Contenido

  1. Overview p. 1
    Introduces the Minichiller 1200w-H chiller and its main features
  2. Spec p. 2
    Lists technical data according to DIN 12876 standard
  3. p. 2 table, 19 rows
  4. p. 1 table, 23 rows

Technical data according to DIN 12876

ParameterValue
max. delivery pressure0,7 bar
max. delivery (suction)18 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
Cooling water connectionG1/2 male
min. cooling water differential pressure3 bar
max. cooling water pressure6 bar
min. filling capacity2,8 l
Volume of expansion2,2 l
Overall dimensions WxDxH **280x490x424 mm
Net weight39 kg
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current12 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S564924

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Minichiller 1200w-H (Order-No.: 3078.0008.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Cooling power at ambient temperature +20°C at 15°C1,2 kW
Cooling power at ambient temperature +20°C at 0°C0,9 kW
Cooling power at ambient temperature +20°C at -10°C0,7 kW
Cooling power at ambient temperature +20°C at -20°C0,35 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,07 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Circulation pumpPressure- and suction pump
max. delivery24 l/min

Pump data for whisper mode: Delivery rate 14 l/min; Delivery pressure 0.2 bar; Delivery rate (suction) 11 l/min; Delivery pressure (suction) 0.18 bar. Reduced sound pressure level 51 dB(A): +/- 4 dB(A).

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires cooling water with a differential pressure of at least 3 bar and no more than 6 bar, so the facility water supply must be verified before installation.

Insufficient or excessive cooling water pressure will prevent proper operation or damage the water-cooled refrigeration circuit.

Source
“min. cooling water differential pressure 3 bar max. cooling water pressure 6 bar”
Installation

The chiller is designed for cooling water intake between 5°C and 20°C, and pipes must be insulated if the water temperature can fall below the dew point.

Cooling water outside this range reduces cooling capacity or causes condensation, affecting reliability and requiring pipe insulation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity is rated at 1.2 kW at 15°C, 0.9 kW at 0°C, 0.7 kW at -10°C and 0.35 kW at -20°C, so the required cooling load must be checked against the actual working temperature.

Cooling capacity drops significantly at lower temperatures, so selecting the unit requires matching the load at the intended set point.

Source
“Cooling power at ambient temperature +20°C at 15°C 1,2 kW at 0°C 0,9 kW at -10°C 0,7 kW at -20°C 0,35 kW”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects handling, ventilation and ignition-source control during installation and service.

A3 flammability classification imposes specific safety requirements on the installation location and maintenance procedures.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant pump option ordered.

Using incompatible media will damage the pump and cause leakage or failure; alternative materials must be specified at purchase.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Installation

The electrical supply must be 208–240 V single or two-phase at 50/60 Hz with a maximum current of 12 A and a 16 A fuse.

The site power connection and protection must be sized to these values to avoid nuisance tripping or inadequate supply.

Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz max. current 12 A Fuse 16 A”
Cumplimiento

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

Unacceptable supply combinations can cause overheating or malfunction, so the site supply quality must be checked against EN60034-1.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit has IP20 protection and must be installed in an environment between 5°C and 40°C ambient temperature.

IP20 offers no water protection and the ambient limits define the allowable installation location and ventilation.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used to avoid corrosion.

Water chemistry must be compatible with these materials to prevent corrosion and fouling of the water-cooled condenser.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

The pump and cooling water connections are M16x1 male and G1/2 male respectively, which determines the hose and fitting selection.

Correct adapters and hoses must be procured before installation to avoid delays and leaks.

Source
“Pump connection M16x1 male Cooling water connection G1/2 male”
Operation

Whisper mode reduces sound pressure to 51 dB(A) ±4 dB(A) but also reduces pump delivery to 14 l/min and 0.2 bar.

If low noise is required, the reduced pump performance in whisper mode must be acceptable for the process.

Source
“Reduced sound pressure level: 51 dB(A): +/- 4 dB(A)”
Selection

The minimum filling capacity is 2.8 l and the expansion volume is 2.2 l, which sets the minimum fluid volume and expansion allowance.

The external loop volume and thermal expansion must be considered to avoid overflow or insufficient fill.

Source
“min. filling capacity 2,8 l Volume of expansion 2,2 l”