Induztr
Resumen técnico Visor de documentos Modelos relacionados

Minichiller 300w-H OLÉ Technical Data

Contenido

  1. Overview p. 1
    Product name and cover page for the Minichiller unit
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 2 table, 5 rows

Technical data according to DIN 12876

ParameterValue
Pressure equipment category Art. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S438237

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires cooling water supply pressure between 3 and 6 bar, so the site water system must be verified before installation.

Insufficient or excessive cooling water pressure will prevent proper operation or damage the water circuit.

Source
“min. cooling water differential pressure 3 bar”
Installation

Cooling water inlet temperature must be kept between 5°C and 20°C, and pipes must be insulated if condensation conditions occur.

Out-of-range cooling water temperature reduces cooling capacity and can cause condensation and corrosion.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops significantly as water temperature falls, from 0.3 kW at 15°C to 0.07 kW at -20°C, so sizing must account for the actual operating point.

Selecting the chiller based only on nominal capacity will underperform at low temperatures.

Source
“at -20°C 0,07 kW”
Safety

The refrigerant is R-290 (propane, A3 flammable), so the installation area must be assessed for flammable refrigerant handling and ventilation.

Flammable refrigerant classification affects site safety requirements and installation approval.

Source
“R-290 (A3, H220)”
Operation

The plastic pump is not resistant to acetone and polyglycols, so these fluids must not be used as heat transfer media.

Using incompatible fluids will damage the pump and void reliable operation.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Proper antifreeze mixture prevents freezing and damage in the cooling circuit at low temperatures.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Site power quality must be checked to avoid invalid supply combinations that could damage the unit.

Source
“-5% voltage and +2% frequency -> not allowed!”
Cumplimiento

The unit is classified as EN55011 Class A Group 1, meaning it is intended for industrial electromagnetic environments, not residential.

Class A equipment may cause interference in residential areas and affects where the unit can be installed.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.

IP20 limits installation to clean, dry environments and rules out washdown or outdoor placement.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must remain between 5°C and 40°C for specified operation.

Exceeding ambient limits reduces cooling performance and may trip alarms or damage components.

Source
“max. ambient temperature 40 °C”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so cooling water must be compatible with these materials.

Incompatible water chemistry will cause corrosion or fouling in the water circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Selection

The unit is water-cooled with a natural refrigerant and very low GWP, which matters for sites with refrigerant regulations.

Water cooling avoids air-cooled condenser heat rejection and the natural refrigerant simplifies compliance.

Source
“Refrigeration machine water-cooled, natural refrigerant”