Contenido

  1. Overview p. 1
    Product introduction and key features of the Minichiller 800w-H
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 2 table, 13 rows
  4. p. 1 table, 23 rows

Technical data according to DIN 12876

ParameterValue
Pump connectionM16x1 male
Cooling water connectionG1/2 male
min. cooling water differential pressure3 bar
max. cooling water pressure6 bar
min. filling capacity2,8 l
Volume of expansion2,2 l
Overall dimensions WxDxH **280x490x424 mm
Power supply requirement208-240V 1~/2~ 50/60Hz
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/23

** Please respect space requirements. See operating conditions at www.huber-online.com

Minichiller 800w-H — Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Cooling power at ambient temperature +20°C at 15°C0,8 kW
Cooling power at ambient temperature +20°C at 0°C0,6 kW
Cooling power at ambient temperature +20°C at -10°C0,45 kW
Cooling power at ambient temperature +20°C at -20°C0,3 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Circulation pumpPressure- and suction pump
max. delivery24 l/min
max. delivery pressure0,7 bar
max. delivery (suction)18 l/min
max. delivery pressure (suction)0,4 bar

Order-No.: 3079.0008.01

Engineering Facts

13 facts · every claim linked to a page
Installation

The chiller requires cooling water with a differential pressure between 3 and 6 bar, so the site water supply must be verified before installation.

Insufficient cooling water differential pressure will prevent the water-cooled refrigeration unit from operating correctly.

Source
“min. cooling water differential pressure 3 bar”
Installation

Cooling water inlet temperature must be kept between 5°C and 20°C, and pipes must be insulated if condensation conditions occur.

Operating outside this cooling water range reduces cooling capacity and risks condensation on uninsulated piping.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Safety

The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, which affects site safety classification and ventilation requirements.

A3 flammability classification requires appropriate handling, ventilation and ignition-source control during installation and service.

Source
“R-290 (A3, H220)”
Operation

The plastic circulation pump is not resistant to acetone or polyglycols, so these media must not be used in the bath circuit.

Using incompatible media will damage the pump; alternative resistant plastic is only available on request at extra cost.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Proper antifreeze mixture prevents freezing and damage when operating near or below 0°C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

The power supply must be checked against this combined tolerance rule to avoid operating the unit outside its specified envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit is rated IP20, so it must be installed in a dry location protected from water ingress.

IP20 offers no water protection, so washdown or wet environments are unsuitable for this chiller.

Source
“Degree of Protection IP20”
Installation

Ambient temperature at the installation site must remain between 5°C and 40°C.

Ambient conditions outside this range affect performance and may void the specified operating envelope.

Source
“min. ambient temperature 5 °C”
Selection

Cooling power drops with lower temperatures, from 0.8 kW at +15°C to 0.3 kW at -20°C, which must be matched to the process heat load.

Sizing must account for reduced cooling capacity at low setpoints to avoid insufficient heat removal.

Source
“at -20°C 0,3 kW”
Performance

Whisper mode reduces pump delivery to 14 l/min and 0.2 bar, which may be insufficient for high-pressure-drop circuits.

If low noise is required, the reduced pump performance must still satisfy the external circuit pressure drop.

Source
“Delivery rate: 14 l/min”
Cumplimiento

The unit is EMC Class A, Group 1, meaning it is intended for industrial environments and may cause interference in residential areas.

Class A equipment may not be suitable for residential or light-commercial installations without additional EMC measures.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.

Planning the electrical connection requires ordering or providing the correct plug for the site supply.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Aggressive or incompatible cooling water can corrode or degrade these materials and cause leaks or fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”