Contenido

  1. Overview p. 1
    Product introduction and main features of Ministat 230
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 14 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
min. filling capacity3 l
Bath capacity with displacement rack1,7 l
Width bath opening WxD / bath depth170 x 85 / 135 mm
sound pressure level +/- 4 dB(A)59 dB(A)
Overall dimensions WxDxH **255x450x476 mm
Net weight37 kg
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current11 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
max. ambient temperature40 °C
min. ambient temperature5 °C
from Serial-No.:S393293

** Please respect space requirements. See operating conditions at www.huber-online.com

Ministat 230 — Technical data according to DIN 12876

CharacteristicValue
Operating temperature range-40...200 °C
Temperature stability at -10°C0,02 K
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Cooling power at 100°C0,42 kW
Cooling power at 20°C0,42 kW
Cooling power at 0°C0,38 kW
Cooling power at -10°C0,33 kW
Cooling power at -20°C0,25 kW
Cooling power at -30°C0,14 kW
Cooling power at -40°C0,05 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,099 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Pressure pumpyes
max. delivery22 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)16 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
max. permissible kin. viscosity50 mm²/s
Bath volume3,5 l

Order-No.: 2015.0005.01

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is a refrigerated heating circulator with a working range of -40 to 200 °C, so it can cover both deep cooling and heating duties in one bath.

Defines whether the circulator can meet the process temperature window before any other sizing is done.

Source
“Operating temperature range -40...200 °C”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, so the installation location must be assessed for A3 refrigerant handling and ignition sources.

A3 classification affects room ventilation, ignition-source control and local refrigerant regulations at the installation site.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Installation

The unit requires a 208-240 V single-phase or two-phase supply with a maximum current of 11 A and a 16 A fuse.

Determines the electrical feed, breaker rating and whether a country-specific plug must be specified at order.

Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz”
Installation

Voltage and frequency tolerances are coupled: ±5 % voltage is only allowed together with ±2 % frequency, and a -5 % voltage with +2 % frequency combination is explicitly not allowed.

Prevents commissioning on a supply whose combined voltage/frequency deviation would exceed the motor and electronics rating.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The enclosure is IP20, so the unit must be installed in a dry, protected location and not exposed to water spray or condensation.

Sets the environmental protection level required at the installation point and rules out washdown or wet areas.

Source
“Degree of Protection IP20”
Installation

The permissible ambient temperature range is 5 to 40 °C, and cooling capacity may drop if the ambient temperature rises above the 20 °C reference.

Room conditions must be checked at the design stage, otherwise the published cooling capacities will not be achieved.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Performance

Cooling capacity falls from 0.42 kW at 20 °C to 0.05 kW at -40 °C, so the low-temperature end has very limited heat removal capability.

Needed to check whether the process heat load can be removed at the lowest required operating temperature.

Source
“at -40°C 0,05 kW”
Performance

The pressure pump delivers up to 22 l/min at 0.7 bar and the suction pump up to 16 l/min at 0.4 bar, with M16x1 male connections.

Used to size external circuit hoses, pressure drop and the achievable flow through the connected application.

Source
“max. delivery 22 l/min”
Operation

The maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s, which limits the choice of heat transfer fluid at low temperatures.

Prevents selecting a fluid that is too viscous at the lowest operating temperature and would overload the pump.

Source
“max. permissible kin. viscosity 50 mm²/s”
Selection

The bath volume is 3.5 l with a minimum filling of 3 l, and it drops to 1.7 l when a displacement rack is used.

Affects fluid charge, heat-up/cool-down time and whether a displacement insert is needed for small external loads.

Source
“Bath volume 3,5 l”
Cumplimiento

The equipment is classified as Class A, Group 1 for electromagnetic disturbance to EN55011, so it is intended for non-domestic environments.

Class A equipment may cause interference in residential areas and must be accounted for in the EMC environment of the lab.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified when ordering.

Avoids a missing or wrong plug on delivery and delays during commissioning.

Source
“with power cable and country-specific plug (please specify when ordering)”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.

The overtemperature cut-out must be set to suit the thermofluid and application to protect the bath and external circuit.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Maintenance

The refrigeration machine is air-cooled, so the condenser needs unobstructed airflow and the stated space requirements must be respected.

Insufficient clearance around an air-cooled condenser reduces cooling capacity and can cause high-pressure trips.

Source
“Please respect space requirements.”