Product introduction and main features of Ministat 230
Specpp. 1-2
Technical data and performance specifications per DIN 12876
p. 2table, 14 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
min. filling capacity
3 l
Bath capacity with displacement rack
1,7 l
Width bath opening WxD / bath depth
170 x 85 / 135 mm
sound pressure level +/- 4 dB(A)
59 dB(A)
Overall dimensions WxDxH **
255x450x476 mm
Net weight
37 kg
Power supply requirement
208-240V 1~/2~ 50/60Hz
max. current
11 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
max. ambient temperature
40 °C
min. ambient temperature
5 °C
from Serial-No.:
S393293
** Please respect space requirements. See operating conditions at www.huber-online.com
Ministat 230 — Technical data according to DIN 12876
Characteristic
Value
Operating temperature range
-40...200 °C
Temperature stability at -10°C
0,02 K
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Cooling power at 100°C
0,42 kW
Cooling power at 20°C
0,42 kW
Cooling power at 0°C
0,38 kW
Cooling power at -10°C
0,33 kW
Cooling power at -20°C
0,25 kW
Cooling power at -30°C
0,14 kW
Cooling power at -40°C
0,05 kW
Refrigeration machine
air-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,099 kg
CO2 equivalent
0,0 t
UN-number
UN 3358
Pressure pump
yes
max. delivery
22 l/min
max. delivery pressure
0,7 bar
Suction pump
yes
max. delivery (suction)
16 l/min
max. delivery pressure (suction)
0,4 bar
Pump connection
M16x1 male
max. permissible kin. viscosity
50 mm²/s
Bath volume
3,5 l
Order-No.: 2015.0005.01
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is a refrigerated heating circulator with a working range of -40 to 200 °C, so it can cover both deep cooling and heating duties in one bath.
Defines whether the circulator can meet the process temperature window before any other sizing is done.
Source
“Operating temperature range -40...200 °C”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, so the installation location must be assessed for A3 refrigerant handling and ignition sources.
A3 classification affects room ventilation, ignition-source control and local refrigerant regulations at the installation site.
Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Installation
The unit requires a 208-240 V single-phase or two-phase supply with a maximum current of 11 A and a 16 A fuse.
Determines the electrical feed, breaker rating and whether a country-specific plug must be specified at order.
Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz”
Installation
Voltage and frequency tolerances are coupled: ±5 % voltage is only allowed together with ±2 % frequency, and a -5 % voltage with +2 % frequency combination is explicitly not allowed.
Prevents commissioning on a supply whose combined voltage/frequency deviation would exceed the motor and electronics rating.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The enclosure is IP20, so the unit must be installed in a dry, protected location and not exposed to water spray or condensation.
Sets the environmental protection level required at the installation point and rules out washdown or wet areas.
Source
“Degree of Protection IP20”
Installation
The permissible ambient temperature range is 5 to 40 °C, and cooling capacity may drop if the ambient temperature rises above the 20 °C reference.
Room conditions must be checked at the design stage, otherwise the published cooling capacities will not be achieved.
Source
“If the ambient temperature rises, the cooling capacity may drop.”
Performance
Cooling capacity falls from 0.42 kW at 20 °C to 0.05 kW at -40 °C, so the low-temperature end has very limited heat removal capability.
Needed to check whether the process heat load can be removed at the lowest required operating temperature.
Source
“at -40°C 0,05 kW”
Performance
The pressure pump delivers up to 22 l/min at 0.7 bar and the suction pump up to 16 l/min at 0.4 bar, with M16x1 male connections.
Used to size external circuit hoses, pressure drop and the achievable flow through the connected application.
Source
“max. delivery 22 l/min”
Operation
The maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s, which limits the choice of heat transfer fluid at low temperatures.
Prevents selecting a fluid that is too viscous at the lowest operating temperature and would overload the pump.
Source
“max. permissible kin. viscosity 50 mm²/s”
Selection
The bath volume is 3.5 l with a minimum filling of 3 l, and it drops to 1.7 l when a displacement rack is used.
Affects fluid charge, heat-up/cool-down time and whether a displacement insert is needed for small external loads.
Source
“Bath volume 3,5 l”
Cumplimiento
The equipment is classified as Class A, Group 1 for electromagnetic disturbance to EN55011, so it is intended for non-domestic environments.
Class A equipment may cause interference in residential areas and must be accounted for in the EMC environment of the lab.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified when ordering.
Avoids a missing or wrong plug on delivery and delays during commissioning.
Source
“with power cable and country-specific plug (please specify when ordering)”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.
The overtemperature cut-out must be set to suit the thermofluid and application to protect the bath and external circuit.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Maintenance
The refrigeration machine is air-cooled, so the condenser needs unobstructed airflow and the stated space requirements must be respected.
Insufficient clearance around an air-cooled condenser reduces cooling capacity and can cause high-pressure trips.