Ministat 240w Circulating Bath Datasheet

Contenido

  1. Overview p. 1
    Product introduction and main features of Ministat 240w
  2. Spec pp. 1-2
    Technical data, performance and dimensions per DIN 12876
  3. p. 2 table, 16 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
min. cooling water differential pressure2 bar
max. cooling water pressure6 bar
Bath volume5,5 l
min. filling capacity3,5 l
Bath capacity with displacement rack2,8 l
Width bath opening WxD / bath depth205 x 85 / 157 mm
Overall dimensions WxDxH **300x465x516 mm
Net weight41 kg
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current11,5 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
max. ambient temperature40 °C
min. ambient temperature5 °C
from Serial-No.:S412173

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Ministat 240w (Order-No.: 2016.0006.01)

CharacteristicValue
Operating temperature range-45...200 °C
Temperature stability at -10°C0,02 K
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at 100°C0,6 kW
Cooling power at 20°C0,6 kW
Cooling power at 0°C0,55 kW
Cooling power at -10°C0,42 kW
Cooling power at -20°C0,35 kW
Cooling power at -30°C0,125 kW
Cooling power at -40°C0,05 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,06 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Pressure pumpyes
max. delivery22 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)16 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 15°C34 l/h

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires cooling water with a differential pressure between 2 and 6 bar, so the facility water supply must be sized accordingly.

Insufficient or excessive cooling water pressure will prevent proper operation or damage the water-cooled condenser.

Source
“min. cooling water differential pressure 2 bar max. cooling water pressure 6 bar”
Installation

Cooling water inlet temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.

Out-of-range cooling water temperature reduces cooling capacity and risks condensation and corrosion.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops significantly at low temperatures, from 0.6 kW at 20 °C to only 0.05 kW at -40 °C.

The user must verify that available cooling capacity at the target low temperature meets the process heat load.

Source
“at -40°C 0,05 kW”
Installation

The electrical supply must be 220-240 V single or two-phase at 50/60 Hz with a 16 A fuse and max current of 11.5 A.

The site electrical installation must provide the correct voltage, phase configuration and protection rating.

Source
“Power supply requirement 220-240V 1~/2~ 50/60Hz max. current 11,5 A Fuse 16 A”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but they must not be combined in opposite directions.

Unstable mains supply outside these combined limits can cause malfunction or damage to the compressor and electronics.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The refrigerant is R-290 (propane, A3 flammable, H220), so the installation area must be assessed for flammable refrigerant handling.

A3 flammable refrigerant imposes ventilation, ignition-source and handling requirements on the installation site.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Operation

The pump delivers up to 22 l/min at 0.7 bar pressure and 16 l/min suction at 0.4 bar, with M16x1 male connections.

External circuit pressure drop and hose sizing must be checked against these pump limits.

Source
“max. delivery 22 l/min max. delivery pressure 0,7 bar”
Operation

The maximum permissible kinematic viscosity of the bath fluid is 50 mm²/s.

Selecting a fluid too viscous at low temperature will overload the pump and reduce circulation.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The bath requires a minimum filling capacity of 3.5 l, and only 2.8 l remains when a displacement rack is used.

Underfilling risks pump cavitation and poor temperature control; the displacement rack reduces usable volume.

Source
“min. filling capacity 3,5 l Bath capacity with displacement rack 2,8 l”
Installation

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient temperature.

IP20 offers no water protection, and ambient limits affect cooling performance and electronics reliability.

Source
“Degree of Protection IP20 max. ambient temperature 40 °C min. ambient temperature 5 °C”
Cumplimiento

The unit is classified as pressure equipment category Art. 4.3 PED and safety class III / FL.

This determines the applicable pressure-equipment compliance and documentation obligations.

Source
“Pressure equipment category Art. 4.3 PED”
Installation

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality may be used.

Aggressive or scaling cooling water will corrode these materials and impair heat transfer.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Operation

The unit consumes 34 l/h of cooling water at 15 °C inlet and 15 °C flow.

The facility must supply adequate cooling water flow rate to sustain rated cooling capacity.

Source
“Consumption at water 15°C, flow 15°C 34 l/h”
Selection

The operating temperature range is -45 to 200 °C with a temperature stability of 0.02 K at -10 °C.

These define whether the unit can meet the process temperature window and stability requirement.

Source
“Operating temperature range -45...200 °C Temperature stability at -10°C 0,02 K”