Contenido

  1. Overview pp. 5-10
    Table of contents, function description, normal operation and shutdown overview
  2. Overview pp. 11-21
    General introduction, safety notes and device overview
  3. Procedure pp. 22-32
    Step-by-step commissioning procedures and initial setup
  4. Overview pp. 33-43
    Functional description of components and operating modes
  5. Procedure pp. 44-61
    Setup mode procedures and parameter configuration
  6. Procedure pp. 62-63
    Operating procedures during normal use
  7. Procedure pp. 64-70
    Interface connections and software update procedures
  8. Procedure pp. 71-78
    Maintenance schedules, service tasks and troubleshooting
  9. Procedure pp. 79-82
    Shutdown procedures and decommissioning steps
  10. pp. 83-84
    Technical annex and temperature range data
  11. p. 9 table, 40 rows
  12. p. 42 table, 21 rows
  13. p. 36 table, 17 rows
  14. p. 34 table, 10 rows
  15. p. 26 table, 6 rows
  16. p. 35 table, 4 rows
  17. p. 70 table, 6 rows
  18. p. 8 table, 40 rows

Содержание (оглавление) страницы 9

РазделНазваниеСтраница
4.2.5Monitoring the Pt100 temperature sensors49
4.2.6Optimum control parameters for optimum temperature control50
4.2.7Sub-category: “Select auto/expert mode”.50
4.2.8Sub-category: “Configuration auto”50
4.2.8.1Sub-category: “Find parameters”50
4.2.8.2Sub-category: “Control Dynamics”52
4.2.8.3Sub-category: “Fluid Properties”53
4.2.8.4Sub-category: “Display parameters”54
4.2.9Sub-category: “Configuration manual”55
4.2.9.1Sub-category: “Change parameters”55
4.2.9.2Sub-category: “Display parameters”56
4.2.9.3Sub-category: “Control structure”56
4.2.10Sub-category: “Reset parameters”56
4.2.11Sub-category: “Display parameters”57
4.2.12Setting the setpoint thresholds57
4.2.13Setting the setpoint58
4.3Filling, venting, degassing and draining58
4.3.1Externally open application58
4.3.1.1Filling and venting externally open application58
4.3.1.2Degassing externally open applications59
4.3.1.3Draining externally open applications60
5Normal operation62
5.1Automatic operation62
5.1.1Temperature control62
5.1.1.1Starting the temperature control process62
5.1.1.2Ending the temperature control process62
5.1.2Temperature control via a created temperature control program63
5.1.2.1Starting the temperature control program63
5.1.2.2Ending/cancelling the temperature control program63
6Interfaces and software update64
6.1Interfaces at the “Pilot ONE®” controller64
6.1.110/100 Mbps Ethernet for RJ45 network sockets64
6.1.2USB-2.0 interface64
6.1.2.1USB-2.0 interface, host65
6.1.2.2USB-2.0 interface, device65
6.2Interfaces on the temperature control unit65
6.2.1Interfaces at the back65
6.2.1.1Connector POKO (floating contact) alarm65
6.2.1.2Jack ECS (External Control Signal) standby66
6.2.1.3Service interface67

На странице приведено оглавление (содержание), а не матрица конфигураций; данные приведены как напечатано.

Restoring the factory settings

Description(A)(B)(C)(D)(E)(F)
AcousticsXXX–––
Display functions Warnings; Messages; Inactive Menu Items; Resolution TemperatureXXX–––
Display modesXXX–––
AutoStart responseXXX–––
Limitations DeltaT limit; max. heating variable; max. cooling variable; max. permissible current consumption (for 230 V temperature control units)XXX–––
Com.G@te (only if Com.G@te is connected) Analog Interface (configuration input) Meaning of AIF input; response on cable break; measurement range settingXXX––X
Calibration–––––X
In case of analog error; current / voltage switching; filter constantXXX––X
Configuration Output AIF output variable; Output range settingXXX––X
Calibration–––––X
RS232/RS485 (Hardware RS, Baud rate, device address); ECS Standby; POKO AlarmXXX–––
Properties (other) Change thermal fluid (thermal fluid, liter indicator, bypass usage); cooling bath / cooling (CC-E only)XXX–––
Sensor calibration––––––
Automated compressorXXX–––
Enter programX––X––
Pump settings Pump speed setpoint; Pump pressure setpoint; Control ModeXXX–––
Fluid properties Thermal fluid; liter indicatorXXX–––
Configuration manual parameters Protective function High Limit Alarm Internal; Low Limit Alarm Internal; High Limit Alarm Process; Low Limit Alarm Process; Hydrostatic Correction; Warning Time Limit (CC-E only); Min. Level (for temperature control units with analogue level sensor); Max. Level (for temperature controllers with analogue level sensor); Freeze Protection (optional)XXX–––
SetpointXXX–––
Setpoint limitation Min. Setpoint; Max. SetpointXXX–––
LanguageXXX–––

“X” = Value is reset, “–” = Value is not reset. (A) = All together; (B) = Device parameters without OT; (C) = Device Parameter; (D) = Program Controller; (E) = Menu; (F) = Com.G@te

Overview of E-grade functions

FunctionE-grade BasicE-grade ExclusiveE-grade Professional
Temperature control
Controller parameterization: predefined1 / TAC2X/––/X–/X
Find parameters: Fast Identification / With Preliminary Test / Estimate Control Parameters–/–/XX/X/–X/X/–
Sensor calibration for control sensor3: x -point255
Monitoring: Low level and overtemperature4XXX
Adjustable alarm thresholdsXXX
VPC5 (Variable Pressure Control)XXX
Ventilation programXXX
Automated compressorXXX
Setpoint limitationXXX
Program controller: x programs / max x steps–/–3/1510/100
Ramp function: linear / non-linear–/–X/–X/X
Temperature control mode: Internal and process–XX
Adjustable max. heating and cooling capacity–XX
Display & operation
Temperature display: 5.7" TouchscreenXXX
Display mode: graphically / numerically large / Explore–/X/––/X/–X/X/–

X = Standard equipment, O = optional, – = not possible. 1 TAC function available as a 30 day evaluation version. 2 True Adaptive Control. 3 Internal Pt100 and External Pt100. 4 For temperature control units with integrated overtemperature protection. 5 For units with variable-speed pump or external bypass.

Table

DesignationSpecification
Calcium carbonate per liter≤ 1.5 mmol/l; corresponds to a water hardness of: ≤ 8.4 °dH (soft)
pH valuebetween 6.0 and 8.5
Ultrapure water, distillatesAdd 0.1 g of sodium carbonate (Na2CO3) per liter
Non-approved waterDistilled, deionized, demineralized, chloric, ferruginous, ammoniacal, contaminated or untreated river water or sea water
Volume circulated (at least)3 l/min. (not valid for cooling baths)
Thermal fluid: Water without ethylene glycol
Useexcluded
Thermal fluid: Water-ethylene glycol mixture
Use-35 °C (with anti-free possible up to -38 °C) to +95 °C
Thermal fluid compositionThe mixture’s temperature must be 10 K below the permissible min. temperature. For the permissible temperature range, refer to the datasheet.→ From page 83, section »Annex«.

Side / Distance in cm (for operation in a tub)

SideAir coolingWater cooling
[A1] TopAir outlet on top of unit: free standing–
[A2] Topcan be located under a benchcan be located under a bench
[B] Leftmin. 20min. 20
[C] Rightmin. 20min. 20
[D] Frontmin. 20min. 20
[E] Rearmin. 20min. 20

Overview of the E-grade variants

Temperature control units / E-gradeE-grade BasicE-grade ExclusiveE-grade Profes-sional
Brewing thermostats–XO
Unistat temperature control units––X
UniCAL––X
Other temperature control unitsXOO

Pin assignment

PinDescriptionSignal
1Current output, T external0 a)/4 - 20 mA or 0 - 10 V If using 0 - 10 V, integrate a 500 Ω resistor.
2Current output, setpoint0 a)/4 - 20 mA or 0 - 10 V
3GND for analog outputsGND
4Analog input (programmable)0 a)/4 - 20 mA or 0 - 10 V Power input: 200 Ω working resistance Voltage input: 100 kΩ input resistance
5Current output, freely programmable0 a)/4 - 20 mA or 0 - 10 V If using 0 - 10 V, integrate a 500 Ω resistor.
6GND for analog inputGND

a) Consult with our Customer Support.

Содержание (оглавление)

РазделНазваниеСтраница
2.10Connecting to the power supply31
2.10.1Connection using socket with protective earth (PE)31
2.10.2Connection via hard wiring32
3Function description33
3.1Function description of the temperature control unit33
3.1.1General functions33
3.1.2Other functions33
3.2Information on the thermal fluids33
3.3To be noted when planning the test34
3.4“Pilot ONE®” controller35
3.4.1Functional overview of “Pilot ONE®”35
3.5Clock/event function38
3.5.1Rechargeable accumulator38
3.5.2Programmable event function38
3.5.2.1Event function “Alarm clock event”38
3.5.2.2Event function “Program event”38
3.6Operation via the touch screen38
3.7Display instruments38
3.7.1The touchscreen [88]39
3.8Control instruments39
3.8.1The touchbuttons39
3.8.2The categories39
3.8.3The sub-categories39
3.8.4The dialogs40
3.9Function examples40
3.9.1Display of software version40
3.9.2Start & Stop40
3.9.3Copying the settings to a data carrier41
3.9.3.1Saving to a USB flash drive41
3.9.3.2Loading from a USB flash drive41
3.9.4Restore factory settings41
3.9.4.1Restore to factory settings without overtemperature protection43
3.9.4.2Restore to factory settings including overtemperature protection43
4Setup mode44
4.1Setup mode44
4.1.1Turning on the temperature control unit44
4.1.2Turning off the temperature control unit44
4.1.3Setting the overtemperature protection45
4.1.3.1General information on the overtemperature protection45
4.1.3.2Setting “OT limit: heating”45

Engineering Facts

13 facts · every claim linked to a page
Safety

The unit must not be installed or started in an ATEX zone without an Ex px cabinet.

Prevents explosion risk and defines the permissible hazardous-area installation scope.

Source
“Do NOT install or start up the temperature control unit within an ATEX zone.”
Installation

For natural-refrigerant units, if the charge exceeds 8 g per m³ of room air, a calibrated gas warning sensor must be installed and maintained every 6–12 months.

Determines room-size limits and mandatory gas detection for NR units.

Source
“Over 8 g refrigerant per m³ room air: A gas warning sensor must be fitted and functioning.”
Installation

Units with more than 1 kg of refrigerant must not be installed below ground level.

Constrains basement or pit installation for larger refrigerant charges.

Source
“Temperature control units with more than 1 kg refrigerant must not be installed below ground level (GL).”
Installation

The power cable must not exceed 3 m so the unit can be disconnected from mains at any time.

Affects electrical installation layout and safe disconnection.

Source
“Do not use a power cable that is longer than 3 m to be able to disconnect the temperature control unit at any time from the mains.”
Installation

For externally open applications, circulation flow must go to the lower connection point and circulation return to the higher point.

Correct piping orientation is required to vent air and operate the application properly.

Source
“the >Circulation flow< [1] connection from the temperature control unit is attached to the lower connection point [B] of the application and the >Circulation return< [2] connection is attached to the higher connection point [A] of the application.”
Installation

Recommended and maximum torques are specified per connection size, e.g. 20 Nm recommended and 24 Nm maximum for M16x1.

Prevents leaks and thread damage when tightening pump connections.

Source
“Recommended torques in Nm”
Operation

The thermal fluid viscosity must not exceed 50 mm²/s at the lowest working temperature and density must not exceed 1 kg/dm³.

Defines fluid selection limits for pump and heat transfer performance.

Source
“The maximum viscosity of the thermal fluid must not exceed 50 mm²/s at the lowest working temperature!”
Operation

Different thermal fluid types must not be mixed; the circuit must be rinsed when changing fluid type.

Prevents property damage and maintains fluid compatibility.

Source
“Do not mix different types of thermofluid (such as mineral oil, silicone oil, synthetic oil, water, etc.) in a thermofluid circuit.”
Operation

Cooling water supply should be shut off even for short downtimes such as overnight.

Reduces risk of condensation and water damage during idle periods.

Source
“Shut off the cooling water supply to the temperature control unit even for shorter downtimes (e.g. overnight).”
Safety

Electrical lines directly below a water-cooled unit must be protected against liquid ingress because condensation can form below the dew point.

Prevents electrical shock hazard from condensation at cooling water connections.

Source
“Electrical lines directly below the temperature control unit must be protected against liquid ingress.”
Maintenance

The air filter mat behind the air inlet connection must be checked and replaced at regular intervals to avoid reduced airflow.

Maintains cooling performance and prevents overheating due to blocked airflow.

Source
“This air filter mat must be checked / replaced at regular intervals so that the air flowing into the temperature control unit is not reduced.”
Cumplimiento

The responsible body must keep F-gas records for at least 5 years and present them to authorities on request.

Defines documentation obligations for units with fluorinated refrigerants.

Source
“The responsible body must keep the records for at least 5 years after their creation and present it to the responsible authority upon request.”
Selection

The unit is classified to DIN 12876 with overheat protection for non-combustible media (NFL) or adjustable overheat/overtemperature plus low-level protection for combustible media (FL).

Determines the safety class and suitable thermal fluid for the application.

Source
“Adjustable overtemperature protection and additional low-level protection”