Installation
The unit requires cooling water with a differential pressure of at least 2 bar and no more than 6 bar at the connection.
To ensure adequate cooling water flow and avoid overpressure damage to the water-cooled condenser.
Source
“min. cooling water differential pressure 2 bar”
Installation
The cooling water intake temperature must be between 5°C and 20°C; below the dew point, cooling water pipes inside the unit must be insulated.
To prevent condensation, corrosion, and ensure rated cooling performance.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
The power supply must be 220-240 V, 1~/2~, 50/60 Hz with a maximum current of 11 A and a 16 A fuse.
To size the electrical supply, protection, and cabling correctly.
Source
“Power supply requirement 220-240V 1~/2~ 50/60Hz”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but a -5% voltage combined with +2% frequency is not allowed.
To avoid operating the compressor and pump outside safe limits and voiding warranty.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The refrigerant is R-290 (propane, A3, H220) with a charge of 0.06 kg, requiring appropriate explosion protection measures.
To assess flammability risk, ventilation requirements, and compliance with local refrigerant regulations.
Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Operation
The unit is a hydraulically sealed circulator with an atmospheric open expansion tank, suitable only for externally closed applications.
To avoid using the unit with open baths or unsealed loops that would cause fluid loss or contamination.
Source
“For externally closed applications.”
Operation
The expansion tank is thermally passive, preventing HTF vapour formation and moisture absorption.
To ensure stable heat transfer fluid properties and reduce maintenance.
Source
“No HTF vapour and no moisture absorption because the expansion tank is thermally passive.”
Performance
Cooling power with ethanol drops from 0.45 kW at 0°C to 0.04 kW at -40°C, so low-temperature operation has very limited heat removal capacity.
To correctly size the cooling load at the lowest required operating temperature.
Source
“at -40°C 0,04 kW”
Performance
The circulation pump delivers a maximum of 25 l/min at 0.9 bar, which limits the achievable flow in high-pressure-drop loops.
To verify that the external circuit pressure drop is compatible with the pump curve.
Source
“max. delivery pressure 0,9 bar”
Operation
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.
To select a suitable heat transfer fluid that remains pumpable at the lowest operating temperature.
Source
“max. permissible kin. viscosity 50 mm²/s”
Installation
The minimum filling capacity is 1.5 l and the expansion tank capacity is 2 l.
To determine the required fluid volume and expansion volume for the external loop.
Source
“min. filling capacity 1,5 l”
Installation
The ambient temperature must be between 5°C and 40°C and the degree of protection is IP20.
To ensure adequate ventilation and avoid overheating or ingress of contaminants.
Source
“max. ambient temperature 40 °C”
Cumplimiento
The equipment is classified as EN55011 Class A, Group 1, so it may cause radio interference in residential areas.
To plan EMC measures and avoid use in sensitive residential or laboratory environments without filtering.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Three-phase devices with current consumption less than 63 A are supplied with cable but without a plug.
To arrange for a suitable plug and electrical connection on site.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”