TC100 Eintauchkühler

Contenido

  1. Overview p. 1
    Introduces the TC100 product and its main features
  2. Spec p. 1
    Technical data table with parameters and values
  3. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range-100...40 °C
Cooling power
at 0°C0,16 kW
at -20°C0,15 kW
at -30°C0,14 kW
at -50°C0,13 kW
at -60°C0,12 kW
at -80°C0,12 kW
at -90°C0,07 kW
at -100°C0,01 kW
Safety classificationI / NFL
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity0,26 kg
CO2 equivalent0,6 t
Refrigerant 2nd stage (ASHRAE, GHS)R-1150 (A3, H220)
Global Warming Potential (GWP)4
Refrigerant quantity 2nd stage0,056 kg
CO2-Äquivalent 2.Stufe0,0 t
Nominal diameter probe48 mm
Length of probe155 mm
Length flexible connection1150 mm
Overall dimensions WxDxH **295x500x570 mm
Net weight61 kg
sound pressure level +/- 4 dB(A)55 dB(A)
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current6,5 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.S411749

** Please respect space requirements. See operating conditions at www.huber-online.com

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit's cooling capacity drops as ambient temperature rises, so it must be sized with margin for warm rooms.

Prevents undersizing when the installation room is warmer than 20 °C.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation

The device must be operated only within an ambient temperature window of 5 to 40 °C.

Defines allowable room conditions for reliable operation and warranty.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

Voltage and frequency tolerances are coupled: ±5 % voltage only with ±2 % frequency, and opposite-sign deviations are not allowed.

Prevents connecting to unstable mains that would violate EN60034-1 tolerances.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Single/two-phase units are delivered with a country-specific plug that must be specified at ordering.

Avoids missing or wrong plug on delivery and rework at site.

Source
“with power cable and country-specific plug (please specify when ordering)”
Installation

Three-phase devices below 63 A come with cable but without plug, so a plug must be sourced separately.

Plans the electrical termination and avoids a missing plug during commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The second refrigerant stage is R-1150 (A3, H220), a flammable refrigerant requiring appropriate handling.

Triggers flammable-refrigerant safety measures for installation and service.

Source
“R-1150 (A3, H220)”
Cumplimiento

The equipment is EMC Class A Group 1, so it may cause interference in residential environments.

Restricts use to industrial/commercial locations and informs EMC planning.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The unit is pressure equipment under Art. 4.3 PED, which affects documentation and conformity obligations.

Determines the regulatory regime applicable to the pressure side.

Source
“Pressure equipment category Art. 4.3 PED”
Installation

Space requirements around the unit must be respected per the operating conditions.

Ensures adequate air flow for the air-cooled condenser and service access.

Source
“Please respect space requirements. See operating conditions”
Selection

The probe has a 48 mm nominal diameter and 155 mm length, which must fit the vessel opening.

Confirms mechanical compatibility with the application vessel before ordering.

Source
“Nominal diameter probe 48 mm Length of probe 155 mm”
Installation

The flexible connection is 1150 mm long, limiting how far the probe can reach from the unit.

Determines the physical layout between the refrigerating unit and the probe.

Source
“Length flexible connection 1150 mm”
Installation

The unit requires a minimum 10 A fuse and permits up to 16 A, with 6.5 A max current.

Sizes the electrical protection and supply circuit correctly.

Source
“max. current 6,5 A min. Fuse 10A max. Fuse 16A”