Selection
The unit's cooling capacity is rated at 0.3 kW at 0 °C, 0.26 kW at -20 °C and 0.2 kW at -30 °C, so it suits low-load immersion cooling only.
Cooling capacity at the actual bath temperature determines whether the unit can hold the process setpoint.
Source
“at 0°C
0,3 kW
at -20°C
0,26 kW
at -30°C
0,2 kW”
Safety
The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, so ignition sources and ventilation must be considered at installation.
Flammable A3 refrigerant drives site safety, ventilation and zoning requirements.
Source
“R-290 (A3, H220)”
Installation
The unit is only rated for ambient temperatures from 5 °C to 40 °C, so the installation room must stay within this band.
Ambient limits define where the air-cooled condenser can be installed and how much heat rejection space is needed.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Performance
Cooling capacity figures are valid at 20 °C room temperature and may drop as ambient temperature rises.
Rated capacity must be derated for warm rooms, otherwise the unit will be undersized.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
Voltage tolerance is ±5 % only with a simultaneous frequency tolerance of ±2 %, and the combination -5 % voltage with +2 % frequency is not allowed.
Supply quality must be checked before connection; some voltage/frequency combinations void the rating.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The power supply requirement is 208-240 V, 1~/2~, 50/60 Hz with max. current 3 A and a fuse between 10 A and 16 A.
Electrical supply and protection sizing must match these values for a compliant connection.
Source
“Power supply requirement
208-240V 1~/2~ 50/60Hz
max. current
3 A
min. Fuse
10A
max. Fuse
16A”
Installation
Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.
The plug type must be stated on the order, otherwise the unit arrives with the wrong connector.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Cumplimiento
The equipment is EMC class A, group 1, so it is intended for industrial environments and may cause interference in residential areas.
Class A equipment restricts where it can be installed and what EMC measures are needed.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Cumplimiento
The unit complies with US-SNAP and applicable EU laws, with the US-SNAP end-use being industrial process refrigeration.
Regulatory end-use classification affects approval and permitted application of the unit.
Source
“This equipment is compliant to US-SNAP and all applicable EU laws. The US-SNAP end-use for this equipment is the industrial process
refrigeration.”
Installation
Space requirements around the unit must be respected as stated in the operating conditions.
Clearances are needed for the air-cooled condenser and for service access.
Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Selection
The flexible condensing probe is 900 mm long with a 13 mm nominal diameter and a 1150 mm flexible connection line.
Probe and line lengths determine whether the unit reaches the vessel and how it is routed.
Source
“Nominal diameter probe
13 mm
Length of probe
900 mm
Length flexible connection
1150 mm”
Installation
The unit weighs 25 kg net with overall dimensions 260x330x415 mm, which sets handling and bench/frame requirements.
Weight and footprint dictate lifting, mounting and space planning.
Source
“Overall dimensions WxDxH **
260x330x415 mm
Net weight
25 kg”
Operation
The refrigeration unit runs continuously, so it should not be switched on/off as a temperature control strategy.
Continuous operation affects energy use, noise and expected compressor life.
Source
“The refrigeration unit works continiously.”
Safety
The sound pressure level is 60 dB(A) ±4 dB(A), which must be considered for personnel exposure and room acoustics.
Noise level matters for workplace limits and installation location.
Source
“sound pressure level +/- 4 dB(A)
60 dB(A)”