Unichiller 015w OLÉ Technical Data

Contenido

  1. Overview p. 1
    Product cover and model identification for Unichiller 015w OLÉ
  2. Spec p. 2
    Technical data according to DIN 12876 for the chiller

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires cooling water with a differential pressure of at least 1 bar and no more than 6 bar, so the site water supply must be verified before installation.

Insufficient differential pressure will prevent proper cooling; excessive pressure may damage the cooling circuit.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is designed for cooling water intake between 5°C and 20°C, and if the water temperature falls below the condensation point, the cooling water pipes inside the unit must be insulated.

Operating outside this range or omitting insulation can cause condensation, corrosion, or reduced performance.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase, so the water supply must be sized for worst-case conditions.

Sizing the cooling water supply only for nominal conditions can lead to insufficient cooling at higher water temperatures.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Using incompatible water chemistry can corrode or degrade wetted materials and void performance expectations.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used; for freeze protection, water mixed with glysantine or ethylene glycol is recommended.

Using incompatible heat transfer fluids will damage the pump and cause downtime.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not allowed.

The electrical supply must be checked against both tolerances simultaneously to avoid invalid operating conditions.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.

Failing to specify the plug type at order can delay commissioning or require field replacement.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation

Three-phase devices with current consumption less than 63A are supplied with cable but without a plug, so a suitable plug must be provided on site.

The installation plan must include a compatible plug and connection for three-phase power.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The refrigerant is R-449A with GHS classification A1, H280, and the charge is 0.19 kg corresponding to 0.3 t CO2 equivalent.

Refrigerant safety classification and CO2 equivalent affect handling, leak reporting, and environmental compliance.

Source
“R-449A (A1, H280)”
Operation

The autostart function restarts the unit after a power failure, which may be undesirable for some processes and should be considered during commissioning.

Automatic restart can cause unexpected cooling or process operation after power restoration.

Source
“Autostart function for power failure”
Installation

The optional Pt100 sensor connection is only available factory fitted and at additional charge, so it must be ordered with the unit.

Retrofitting the external Pt100 display input is not possible in the field.

Source
“only available factory fitted, additional charge”
Installation

The unit has IP20 protection and must be installed in an environment with ambient temperature between 5°C and 40°C.

IP20 offers no water protection, and ambient limits define the acceptable installation location.

Source
“min. ambient temperature 5 °C”