Product name and model identification for the chiller unit
Specp. 2
Technical data table according to DIN 12876 standard
p. 2table, 6 rows
Technical data according to DIN 12876
Parameter
Value
Power supply requirement
208V 2~ 60Hz
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/21
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller requires cooling water with a minimum differential pressure of 1 bar and a maximum supply pressure of 6 bar.
Ensures the water circuit can deliver adequate flow without overpressurizing the unit.
Source
“min. cooling water differential pressure 1 bar”
Installation
Cooling water inlet temperature must be between 5°C and 20°C; below the condensation point the cooling water pipes inside the unit must be insulated.
Prevents condensation and ensures rated cooling performance.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.
Affects sizing of the cooling water supply and expected process capacity.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water should be used.
Guides water quality and material compatibility to avoid corrosion or fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%; combinations such as -5% voltage with +2% frequency are not allowed.
Defines acceptable power supply quality for reliable operation and warranty.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Single/two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.
Avoids delays and ensures the correct plug for the installation site.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety
The unit is classified IP20, so it must be installed in a dry environment protected from water ingress.
Prevents electrical hazards and ensures compliance with installation environment requirements.
Source
“Degree of Protection IP20”
Operation
Ambient temperature must remain between 5°C and 40°C for specified operation.
Operating outside this range may affect performance, reliability and safety.
Source
“min. ambient temperature 5 °C”
Selection
The refrigerant R-449A has a GWP of 1397, which matters for environmental compliance and potential refrigerant regulations.
Influences environmental permitting and total cost of ownership.
Source
“Global Warming Potential (GWP) 1397”
Installation
The unit requires a minimum filling capacity of 2.5 l and has 4 l expansion volume.
Ensures the external circuit has adequate heat transfer fluid volume for stable operation.
Source
“min. filling capacity 2,5 l”
Performance
At -10°C the cooling power is only 0.4 kW, so process loads must be sized accordingly at low temperatures.
Prevents oversizing expectations and ensures adequate cooling at the lowest operating temperature.
Source
“at -10°C 0,4 kW”
Operation
The circulation pump delivers 30 l/min maximum and 3.8 bar maximum pressure, with flow decreasing as pressure rises.
Helps match the pump to the external circuit pressure drop and required flow.