Contenido

  1. Overview p. 1
    Product name and model identification on cover
  2. Spec p. 2
    Technical data according to DIN 12876 standard

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -10 to 40 °C, so process requirements outside this window cannot be met.

Defines the application envelope and excludes processes needing lower or higher temperatures.

Source
“Operating temperature range -10...40 °C”
Selection

Cooling capacity drops from 2.2 kW at 15 °C to 1 kW at -10 °C, so the unit must be sized at the actual operating temperature.

Prevents undersizing when the process runs at low temperatures.

Source
“at 15°C 2,2 kW at 0°C 1,6 kW at -10°C 1 kW”
Safety

The refrigerant is R-290 (A3, H220) with a charge of 0.25 kg, which is flammable and requires appropriate site precautions.

Flammable refrigerant classification affects room ventilation, ignition sources and safety compliance.

Source
“R-290 (A3, H220)”
Installation

The unit requires a 220-240 V 1~ 50 Hz supply with a 16 A fuse and draws up to 6.5 A.

Determines the electrical circuit, breaker rating and plug/cable configuration at the installation site.

Source
“Power supply requirement 220-240V 1~ 50Hz max. current 6,5 A Fuse 16 A”
Installation

The enclosure is IP20, so it must be installed in a dry indoor location protected from water ingress.

IP20 is not water-protected and restricts placement in washdown or outdoor areas.

Source
“Degree of Protection IP20”
Installation

Ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may fall as ambient rises.

Room conditions at the installation site directly affect achievable cooling performance.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and combining -5% voltage with +2% frequency is not allowed.

Defines acceptable mains quality and warns against marginal supply conditions that could damage the unit.

Source
“-5% voltage and +2% frequency -> not allowed!”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used as the heat transfer medium.

Prevents chemical attack and pump failure when selecting the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Guides correct fluid preparation to avoid freezing and damage at low temperatures.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.

Ensures the correct plug type is supplied for the destination country and avoids rework on site.

Source
“with power cable and country-specific plug (please specify when ordering)”
Maintenance

A drain valve is available as an optional accessory, which simplifies fluid changes and draining during maintenance.

Planning the drain valve at purchase reduces effort for fluid replacement and service.

Source
“Optional accessories: drain valve #6839”
Performance

The circulation pump delivers 21 l/min at 0.5 bar and up to 29 l/min maximum with a maximum delivery pressure of 1 bar.

Needed to check whether the pump can overcome the pressure drop of the external circuit.

Source
“at 0,5 bar 21 l/min max. delivery 29 l/min max. delivery pressure 1 bar”
Installation

The pump connection is G3/4 male and a matching hose coupling is included, so external piping must suit this thread.

Determines the fittings and hoses required for connecting the process circuit.

Source
“Pump connection G3/4 male”
Selection

The minimum filling capacity is 3.8 l and the expansion tank is 1.7 l, which sets the minimum fluid volume and expansion allowance.

Affects the amount of heat transfer fluid to procure and the allowable thermal expansion in the loop.

Source
“min. filling capacity 3,8 l expansion tank 1,7 l”