Introduces the Unichiller 025T model and its main features
Specpp. 1-2
Lists performance, temperature, and physical specifications per DIN 12876
p. 2table, 3 rows
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S591844
Technical data according to DIN 12876 — Unichiller 025T (Order-No.: 3054.0012.01)
Parameter
Value
Operating temperature range
-10...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Temperature stability at -10°C
0,5 K
Safety classification
I / NFL
Cooling power at 15°C
2,5 kW
Cooling power at 0°C
1,2 kW
Cooling power at -10°C
0,6 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
1,2 kg
CO2 equivalent
1,7 t
UN-number
UN 2857
Circulation pump:
B
max. delivery
25 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
20 l/min
Delivery at 1,0 bar
17 l/min
Delivery at 2,0 bar
10 l/min
Delivery at 3,0 bar
5 l/min
Pump connection
G3/4 male
min. filling capacity
2,5 l
Volume of expansion
2 l
Overall dimensions WxDxH **
450x510x1230 mm
Net weight
144 kg
sound pressure level +/- 4 dB(A)
74 dB(A)
Power supply requirement
230V 1~ 50Hz
max. current
11 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The chiller is rated for an operating temperature range of -10 to 40 °C, so process requirements outside this window cannot be met.
Defines the process temperature envelope for which the unit can be selected.
Source
“Operating temperature range
-10...40 °C”
Performance
Cooling capacity drops from 2.5 kW at 15 °C to 0.6 kW at -10 °C, so the required heat load must be checked at the actual operating temperature.
Prevents oversizing assumptions based on a single capacity figure.
Source
“Cooling power
at 15°C
2,5 kW
at 0°C
1,2 kW
at -10°C
0,6 kW”
Installation
The unit requires a 230 V single-phase 50 Hz supply with a 16 A fuse and draws up to 11 A, so the electrical installation must be sized accordingly.
Determines the site power supply, cable and protection device rating.
Source
“Power supply requirement
230V 1~ 50Hz
max. current
11 A
Fuse
16 A”
Installation
The chiller is air-cooled and its cooling capacity is valid at 20 °C room temperature, so adequate ventilation and ambient conditions must be provided.
Ambient temperature directly affects achievable cooling performance and must be considered during siting.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The permitted ambient temperature range is 5 to 40 °C, so the installation location must stay within these limits.
Defines the acceptable room conditions for reliable operation.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
Voltage and frequency tolerances are ±5 % and ±2 % respectively, but not simultaneously in the same direction, so supply quality must be verified.
Avoids invalid supply combinations that could damage the unit or void compliance.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety
The refrigerant is R-449A with GHS classification A1/H280 and GWP 1397, so leak detection and refrigerant handling procedures are required.
Affects safety measures, environmental reporting and service planning.
Source
“Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)”
Cumplimiento
The unit is classified as pressure equipment category Art. 4.3 PED, so its installation must respect the applicable pressure equipment directive requirements.
Determines the regulatory framework for installation and commissioning.
Source
“Pressure equipment category
Art. 4.3 PED”
Installation
The pump connection is G3/4 male, so mating hoses and fittings must match this thread.
Ensures correct hydraulic connection without adapters.
Source
“Pump connection
G3/4 male”
Operation
The minimum filling capacity is 2.5 l and expansion volume is 2 l, so the external circuit must be filled and sized within these limits.
Prevents underfilling or overfilling of the expansion system.
Source
“min. filling capacity
2,5 l
Volume of expansion
2 l”
Installation
The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.
Restricts the installation environment and enclosure requirements.
Source
“Degree of Protection
IP20”
Operation
The sound pressure level is 74 dB(A) ±4 dB(A), so noise exposure and acoustic planning must be considered at the installation site.
Affects workplace noise limits and location selection.
Source
“sound pressure level +/- 4 dB(A)
74 dB(A)”
Cumplimiento
The equipment is EMC class A group 1 to EN55011, so it is intended for industrial environments and may cause interference in residential areas.
Determines suitability for industrial versus residential installation.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single/two-phase devices are delivered with a country-specific plug that must be specified at ordering, so the correct plug type must be confirmed.
Avoids delivery of an incompatible mains plug.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”