Introduces the Unichiller 025Tw model and its main features
Specpp. 1-2
Lists performance, temperature, and dimensional specifications per DIN 12876
p. 2table, 8 rows
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
Power supply requirement
208V 2~ 60Hz
max. current
13 A
Fuse
2x16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S542187
Technical data according to DIN 12876 — Unichiller 025Tw (Order-No.: 3054.0018.01)
Parameter
Value
Operating temperature range
-10...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at 15°C
2,5 kW
Cooling power at 0°C
1,2 kW
Cooling power at -10°C
0,6 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,4 kg
CO2 equivalent
0,6 t
Circulation pump
B
max. delivery
30 l/min
max. delivery pressure
3,8 bar
Delivery at 0,12 bar
29 l/min
Delivery at 0,5 bar
25 l/min
Delivery at 1,0 bar
21 l/min
Delivery at 1,5 bar
17 l/min
Delivery at 2,0 bar
13,5 l/min
Delivery at 2,5 bar
10 l/min
Delivery at 3,0 bar
6,5 l/min
Pump connection
G3/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
120 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
2,5 l
Volume of expansion
4 l
Overall dimensions WxDxH **
400x440x1230 mm
Net weight
138 kg
** см. сноску в оригинале (Overall dimensions WxDxH **)
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller requires cooling water with a differential pressure of at least 1 bar and no more than 6 bar at the inlet.
Ensures the water supply system can provide adequate flow without exceeding pressure limits.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit is designed for cooling water intake temperatures between 5°C and 20°C, and pipes must be insulated if below the condensation point.
Prevents condensation and ensures specified cooling performance.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, and flow rate may need to increase.
Affects sizing of the cooling water supply and expected performance.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, so only compatible water quality should be used.
Prevents corrosion and damage to the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento
Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, but combinations like -5% voltage and +2% frequency are not allowed.
Ensures the power supply meets the unit's electrical requirements.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Single/two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.
Avoids delays and ensures correct plug type for the installation site.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety
The refrigerant R-449A has a Global Warming Potential of 1397 and a CO2 equivalent of 0.6 t, which must be considered for environmental regulations.
Affects compliance with refrigerant regulations and leak reporting requirements.
Source
“Global Warming Potential (GWP) 1397”
Operation
The unit has a safety classification I / NFL according to DIN 12876.
Indicates the safety level for use with flammable or non-flammable fluids.
Source
“Safety classification I / NFL”
Installation
The minimum filling capacity is 2.5 liters and the expansion volume is 4 liters.
Ensures proper fluid volume for safe operation and expansion.
Source
“min. filling capacity 2,5 l”
Installation
The pump connection is G3/4 male and the cooling water connection is G1/2 male.
Needed to select matching fittings and hoses for installation.
Source
“Pump connection G3/4 male”
Operation
The unit is for externally closed applications only.
Prevents misuse in open systems that could cause contamination or overflow.
Source
“For externally closed applications.”
Installation
The maximum ambient temperature is 40°C and minimum is 5°C.
Ensures the installation location meets environmental conditions for reliable operation.