Technical data according to DIN 12876
Parameter
Value
min. filling capacity
2,5 l
Volume of expansion
4 l
Overall dimensions WxDxH **
400x440x1230 mm
Net weight
141 kg
sound pressure level +/- 4 dB(A)
61 dB(A)
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
8,5 A
Fuse (3 phase)
3x10A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S340827
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unichiller 030Tw-H (Order-No.: 3025.0052.01)
Parameter
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at 15°C
3 kW
Cooling power at 0°C
2,75 kW
Cooling power at -10°C
2 kW
Cooling power at -20°C
1 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,65 kg
CO2 equivalent
0,9 t
Circulation pump
B
max. delivery
26 l/min
max. delivery pressure
3 bar
Delivery at 0,2 bar
22 l/min
Delivery at 0,5 bar
20 l/min
Delivery at 1,0 bar
16 l/min
Delivery at 1,5 bar
12 l/min
Delivery at 2,0 bar
8 l/min
Pump connection
G3/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
99 l/h
Consumption at water 15°C, flow 0°C
99 l/h
Consumption at water 15°C, flow -10°C
65 l/h
Consumption at water 15°C, flow -20°C
25 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires a cooling water supply with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar.
Ensures the site cooling water loop can provide adequate flow and pressure without overpressurizing the unit.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit must be operated with cooling water intake between 5°C and 20°C, and pipes must be insulated if water temperature can fall below the condensation point.
Defines the allowable cooling water temperature window and condensation prevention measures for reliable installation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity drops as cooling water temperature rises, and cooling water flow rate may need to increase accordingly.
Affects sizing of the cooling water supply and expected performance at the installation site.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and material compatibility to avoid corrosion or damage to the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
Three-phase devices with current consumption less than 63 A are delivered with a cable but without a plug, so the customer must provide a suitable plug and connection.
Clarifies electrical supply scope and what the installer must provide for commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Cumplimiento
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
Prevents installation on a supply where voltage and frequency deviations combine outside the permitted envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, and the unit contains 0.65 kg with a CO2 equivalent of 0.9 t.
Informs refrigerant safety classification, charge quantity and CO2 equivalent for site safety and environmental compliance.
Source
“R-449A (A1, H280)”
Operation
The unit has a minimum filling capacity of 2.5 l and an expansion volume of 4 l, which must be respected when filling the bath.
Ensures correct fill level for safe operation and proper expansion tank function.
Source
“min. filling capacity 2,5 l”
Installation
The unit has protection degree IP20 and must be installed in an ambient temperature between 5°C and 40°C.
Defines environmental limits and ingress protection for the installation location.
Source
“max. ambient temperature 40 °C”
Installation
The unit requires a 400 V 3~ 50 Hz three-phase supply with a maximum current of 8.5 A and 3x10 A fusing.
Specifies the electrical supply and protection required for safe connection.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Performance
The circulation pump delivers 26 l/min maximum and 3 bar maximum delivery pressure, with flow decreasing as pressure rises.
Allows checking whether the pump can overcome the pressure drop of the external application circuit.
Source
“max. delivery pressure 3 bar”
Cumplimiento
The equipment is classified for electromagnetic disturbance as EN55011 Class A, Group 1.
Indicates EMC class for installation environment and potential need for filtering or separation from sensitive equipment.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”