Installation
The chiller requires a three-phase 400 V supply with a 3x10 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply, cable and protection device rating before installation.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
The unit is water-cooled and needs a cooling water supply with at least 1 bar differential pressure and no more than 6 bar.
Cooling water pressure must be verified on site to ensure sufficient flow and avoid overpressure damage.
Source
“min. cooling water differential pressure 1 bar”
Installation
Cooling water intake temperature must be between 5 °C and 20 °C, and pipes must be insulated if below the condensation point.
Defines the permissible cooling water temperature range and condensation protection requirements.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase.
Performance must be derated if the available cooling water is warmer than the rated 15 °C.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Prevents corrosion and material damage by selecting suitable cooling water quality.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Ensures the site power quality is within the allowed envelope before commissioning.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices below 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.
Affects electrical installation planning and the need to procure a matching plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The unit is rated IP20 and must be operated in ambient temperatures from 5 °C to 40 °C.
Defines the acceptable operating environment and protection against ingress.
Source
“min. ambient temperature 5 °C”
Safety
The refrigerant is R-449A with GHS classification H280 and a GWP of 1397, which matters for handling and leak reporting.
Informs refrigerant safety classification and environmental reporting obligations.
Source
“R-449A (A1, H280)”
Selection
The chiller is intended for externally closed applications, which constrains how it can be integrated into the process.
Confirms the hydraulic configuration required for correct use.
Source
“For externally closed applications.”
Maintenance
The expansion tank is atmospheric open and the unit has a digital level indicator, so level monitoring is part of routine checks.
Indicates maintenance attention points for the expansion tank and fluid level.
Source
“Housing, atmospheric open expansion tank and copper soldered evaporator made of stainless steel. With digital level indicator.”
Installation
Space requirements around the unit must be respected according to the operating conditions published online.
Prevents installation in locations without adequate clearance for operation and service.
Source
“Please respect space requirements. See operating conditions at www.huber-online.com”