Unichiller 045T-H Technical Datasheet

Contenido

  1. Overview p. 1
    Introduces the Unichiller 045T-H chiller unit and its main features
  2. Spec pp. 1-2
    Performance, temperature, cooling capacity and operating data per DIN 12876
  3. Spec p. 2
    Physical dimensions, weight and process connection details
  4. p. 2 table, 4 rows
  5. p. 1 table, 35 rows

Technical data according to DIN 12876

ParameterValue
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S605589

Technical data according to DIN 12876

ParameterValue
Order-No.3055.0001.01
ModelUnichiller 045T-H
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,2 K
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C4,5 kW
Cooling power at 0°C4 kW
Cooling power at -10°C2,7 kW
Cooling power at -20°C1,4 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity2,55 kg
CO2 equivalent3,6 t
Circulation pumpB
max. delivery26 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar20 l/min
Delivery at 1,0 bar16 l/min
Delivery at 1,5 bar12 l/min
Delivery at 2,0 bar8 l/min
Pump connectionG3/4 male
min. filling capacity3,5 l
Volume of expansion3,5 l
Overall dimensions WxDxH **500x552x1451 mm
Net weight187 kg
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)10 A
Fuse (3 phase)3x10A
Pressure equipment categoryArt. 4.3 PED

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller requires a three-phase 400 V supply with a 3x10 A fuse, so the site must provide a suitable three-phase feed and protection.

Electrical supply and fuse rating must be confirmed before installation to avoid rework or undersized protection.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with a cable but without a plug, so the installer must provide and fit the site plug.

This affects installation planning and the bill of materials for the electrical connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit is IP20, so it must be installed in a dry indoor location protected from water and dust ingress.

IP20 limits where the chiller can be placed and what environmental protection is needed.

Source
“Degree of Protection IP20”
Operation

The chiller may only be operated between 5 °C and 40 °C ambient temperature, so the room climate must be kept within this range.

Ambient temperature outside this range can affect cooling performance and safe operation.

Source
“min. ambient temperature 5 °C”
Performance

Cooling capacity is rated at 20 °C room temperature and may drop if the ambient temperature rises, so selection must account for the actual site ambient.

Sizing the chiller on catalogue capacity without ambient correction can lead to insufficient cooling at the real installation site.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

Voltage and frequency tolerances must occur together within ±5 % and ±2 %, and certain combinations such as -5 % voltage with +2 % frequency are not allowed.

The site power quality must be checked against these combined tolerances to avoid invalid supply conditions.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The refrigerant is R-449A with GHS classification H280 and a GWP of 1397, so refrigerant handling and leak response must follow the applicable safety rules.

Refrigerant type and hazard classification affect safety planning, service procedures and environmental compliance.

Source
“R-449A (A1, H280)”
Cumplimiento

The CO2 equivalent of the refrigerant charge is 3.6 t, which may trigger refrigerant leak-check and record-keeping obligations.

Operators need to know the CO2 equivalent to determine applicable refrigerant regulations and maintenance obligations.

Source
“CO2 equivalent 3,6 t”
Installation

The pump connection is G3/4 male, so the external circuit must be prepared with matching G3/4 female fittings.

Knowing the connection thread avoids mismatched fittings and delays during commissioning.

Source
“Pump connection G3/4 male”
Operation

The minimum filling capacity is 3.5 l, so the external circuit and expansion volume must be designed to hold at least this volume.

Underfilling can prevent proper circulation and temperature control.

Source
“min. filling capacity 3,5 l”
Installation

The unit is air-cooled and requires space for air flow, so the stated space requirements must be respected at the installation site.

Insufficient clearance can reduce cooling performance and cause overheating.

Source
“Please respect space requirements.”
Cumplimiento

The equipment is classified as EN55011 Class A Group 1, so it is intended for industrial electromagnetic environments rather than residential ones.

This affects where the chiller can be installed and whether additional EMC measures may be needed.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The unit has adjustable overtemperature protection according to DIN 12876, which must be set correctly for the application.

Correct setting of the overtemperature protection is essential for safe operation of the external circuit.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The chiller is intended for externally closed applications, so the external circuit must be designed as a closed loop.

Using it with an open external circuit would be outside the intended application.

Source
“For externally closed applications.”