Introduces the Unichiller 045T-H chiller unit and its main features
Specpp. 1-2
Performance, temperature, cooling capacity and operating data per DIN 12876
Specp. 2
Physical dimensions, weight and process connection details
p. 2table, 4 rows
p. 1table, 35 rows
Technical data according to DIN 12876
Parameter
Value
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S605589
Technical data according to DIN 12876
Parameter
Value
Order-No.
3055.0001.01
Model
Unichiller 045T-H
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Temperature stability at -10°C
0,2 K
Safety classification
III / FL
Heating power
2 kW
Cooling power at 15°C
4,5 kW
Cooling power at 0°C
4 kW
Cooling power at -10°C
2,7 kW
Cooling power at -20°C
1,4 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
2,55 kg
CO2 equivalent
3,6 t
Circulation pump
B
max. delivery
26 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
20 l/min
Delivery at 1,0 bar
16 l/min
Delivery at 1,5 bar
12 l/min
Delivery at 2,0 bar
8 l/min
Pump connection
G3/4 male
min. filling capacity
3,5 l
Volume of expansion
3,5 l
Overall dimensions WxDxH **
500x552x1451 mm
Net weight
187 kg
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
10 A
Fuse (3 phase)
3x10A
Pressure equipment category
Art. 4.3 PED
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller requires a three-phase 400 V supply with a 3x10 A fuse, so the site must provide a suitable three-phase feed and protection.
Electrical supply and fuse rating must be confirmed before installation to avoid rework or undersized protection.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
Three-phase devices below 63 A are delivered with a cable but without a plug, so the installer must provide and fit the site plug.
This affects installation planning and the bill of materials for the electrical connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The unit is IP20, so it must be installed in a dry indoor location protected from water and dust ingress.
IP20 limits where the chiller can be placed and what environmental protection is needed.
Source
“Degree of Protection IP20”
Operation
The chiller may only be operated between 5 °C and 40 °C ambient temperature, so the room climate must be kept within this range.
Ambient temperature outside this range can affect cooling performance and safe operation.
Source
“min. ambient temperature 5 °C”
Performance
Cooling capacity is rated at 20 °C room temperature and may drop if the ambient temperature rises, so selection must account for the actual site ambient.
Sizing the chiller on catalogue capacity without ambient correction can lead to insufficient cooling at the real installation site.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
Voltage and frequency tolerances must occur together within ±5 % and ±2 %, and certain combinations such as -5 % voltage with +2 % frequency are not allowed.
The site power quality must be checked against these combined tolerances to avoid invalid supply conditions.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The refrigerant is R-449A with GHS classification H280 and a GWP of 1397, so refrigerant handling and leak response must follow the applicable safety rules.
Refrigerant type and hazard classification affect safety planning, service procedures and environmental compliance.
Source
“R-449A (A1, H280)”
Cumplimiento
The CO2 equivalent of the refrigerant charge is 3.6 t, which may trigger refrigerant leak-check and record-keeping obligations.
Operators need to know the CO2 equivalent to determine applicable refrigerant regulations and maintenance obligations.
Source
“CO2 equivalent 3,6 t”
Installation
The pump connection is G3/4 male, so the external circuit must be prepared with matching G3/4 female fittings.
Knowing the connection thread avoids mismatched fittings and delays during commissioning.
Source
“Pump connection G3/4 male”
Operation
The minimum filling capacity is 3.5 l, so the external circuit and expansion volume must be designed to hold at least this volume.
Underfilling can prevent proper circulation and temperature control.
Source
“min. filling capacity 3,5 l”
Installation
The unit is air-cooled and requires space for air flow, so the stated space requirements must be respected at the installation site.
Insufficient clearance can reduce cooling performance and cause overheating.
Source
“Please respect space requirements.”
Cumplimiento
The equipment is classified as EN55011 Class A Group 1, so it is intended for industrial electromagnetic environments rather than residential ones.
This affects where the chiller can be installed and whether additional EMC measures may be needed.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety
The unit has adjustable overtemperature protection according to DIN 12876, which must be set correctly for the application.
Correct setting of the overtemperature protection is essential for safe operation of the external circuit.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation
The chiller is intended for externally closed applications, so the external circuit must be designed as a closed loop.
Using it with an open external circuit would be outside the intended application.