Introduces the Unichiller 060Tw model and its main features
Specpp. 1-2
Lists performance, temperature, and dimensional specifications per DIN 12876
p. 2table, 14 rows
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
max. cooling water pressure
6 bar
min. filling capacity
5,9 l
Volume of expansion
7,5 l
Overall dimensions WxDxH **
600x600x1450 mm
Net weight
201 kg
sound pressure level +/- 4 dB(A)
57 dB(A)
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
10 A
Fuse (3 phase)
3x16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S456818
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unichiller 060Tw (Order-No.: 3026.0106.01)
Parameter
Value
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at 15°C
6 kW
Cooling power at 0°C
5 kW
Cooling power at -10°C
3,1 kW
Cooling power at -20°C
1,7 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
1,15 kg
CO2 equivalent
1,6 t
Circulation pump
C3
max. delivery
80 l/min
max. delivery pressure
5,6 bar
Delivery at 0,35 bar
79 l/min
Delivery at 0,5 bar
76 l/min
Delivery at 1,0 bar
71 l/min
Delivery at 1,5 bar
66 l/min
Delivery at 2,0 bar
61 l/min
Delivery at 3,0 bar
49 l/min
Delivery at 4,0 bar
35 l/min
Delivery at 5,0 bar
16 l/min
Pump connection
G1 1/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
330 l/h
Consumption at water 15°C, flow 0°C
230 l/h
Consumption at water 15°C, flow -10°C
152 l/h
Consumption at water 15°C, flow -20°C
70 l/h
min. cooling water differential pressure
1 bar
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires a minimum cooling water differential pressure of 1 bar to operate.
Ensures the site's cooling water supply can provide adequate flow and pressure before installation.
Source
“min. cooling water differential pressure
1 bar”
Installation
The unit must be supplied with cooling water between 5°C and 20°C, and pipes must be insulated if water temperature can fall below the condensation point.
Prevents condensation and corrosion damage; defines the acceptable cooling water supply conditions.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling power ratings are valid only at 20°C room temperature, 15°C cooling water inlet, and 1 bar differential pressure.
Allows correct sizing; actual cooling capacity will differ if site conditions deviate from these reference values.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Performance
Higher cooling water temperature reduces cooling power and may require increased cooling water flow rate.
Engineers must account for derating and higher water consumption when the available cooling water is warmer than 15°C.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Prevents corrosion and fouling; guides water quality and material compatibility checks.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Installation
Three-phase devices with current consumption below 63 A are delivered with a cable but without a plug.
The installer must provide and fit a suitable plug or hard-wire the connection on site.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Cumplimiento
Voltage and frequency tolerances are ±5% and ±2% respectively, but opposite-sign deviations (e.g., -5% voltage with +2% frequency) are not allowed.
Defines the acceptable power supply quality; out-of-tolerance supply can damage the compressor and electronics.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, and the unit contains 1.15 kg (1.6 t CO2 equivalent).
Informs refrigerant handling, leak detection, and F-gas reporting obligations.
Source
“R-449A (A1, H280)”
Installation
The unit has protection degree IP20 and must be installed in an environment between 5°C and 40°C ambient temperature.
IP20 offers no water protection, and ambient limits define the allowable installation location.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
The unit requires a 400 V 3~ 50 Hz supply with a 3x16 A fuse and maximum current of 10 A.
Determines the electrical supply, protection device rating, and cabling required at the installation site.
Source
“Power supply (3 Phase)
400V 3~ 50Hz”
Operation
The expansion tank is atmospheric open and the unit is intended for externally closed applications.
Prevents pressurizing the expansion tank and guides correct connection of the external circuit.
Source
“Housing, atmospheric open expansion tank
and copper soldered evaporator made of stainless steel.”
Maintenance
The digital level indicator helps monitor the fill level; minimum filling capacity is 5.9 l and expansion volume is 7.5 l.
Supports correct filling and periodic level checks to avoid pump cavitation or overflow.