Unichiller 070Tw Technical Datasheet

Contenido

  1. Overview p. 1
    Introduces the Unichiller 070Tw chiller unit and its main features
  2. Spec pp. 1-2
    Lists performance, temperature, and physical specifications per DIN 12876
  3. p. 2 table, 15 rows
  4. p. 1 table, 35 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow -10°C264 l/h
min. cooling water differential pressure1,5 bar
max. cooling water pressure6 bar
min. filling capacity5,9 l
Volume of expansion7,5 l
Overall dimensions WxDxH **600x600x1450 mm
Net weight206 kg
Power supply (3 Phase)400V 3~ 50Hz
max. current9 A
Fuse (3 phase)3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S433366

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

CharacteristicUnichiller 070Tw
Order-No.3016.0030.01
Operating temperature range-10...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at 15°C7 kW
Cooling power at 0°C4,2 kW
Cooling power at -10°C2,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,35 kg
CO2 equivalent1,9 t
Circulation pumpC3
max. delivery84 l/min
max. delivery pressure5,6 bar
Delivery at 0,33 bar81 l/min
Delivery at 0,5 bar79 l/min
Delivery at 1,0 bar74 l/min
Delivery at 1,5 bar69 l/min
Delivery at 2,0 bar64 l/min
Delivery at 2,5 bar58 l/min
Delivery at 3,0 bar52 l/min
Delivery at 3,5 bar46 l/min
Delivery at 4,0 bar39 l/min
Delivery at 4,5 bar30 l/min
Delivery at 5,0 bar21 l/min
Delivery at 5,5 bar7 l/min
Pump connectionG1 1/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 15°C474 l/h
Consumption at water 15°C, flow 0°C230 l/h

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires cooling water inlet temperature between 5°C and 20°C, and if it drops below the condensation point the cooling water pipes inside the unit must be insulated.

Defines the permissible cooling water supply conditions and mandatory insulation to prevent condensation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power ratings are valid only at 20°C room temperature, 15°C cooling water inlet and 1.5 bar differential pressure, so actual capacity will differ under other conditions.

Prevents overestimating cooling capacity when site conditions deviate from the reference point.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1,5 bar differential pressure between cooling water inlet and -outlet.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and material compatibility to avoid corrosion or damage.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed.

Defines acceptable power supply quality for reliable operation and warranty compliance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices with current consumption less than 63A are delivered with cable but without plug, so the customer must provide a suitable plug.

Affects electrical installation planning and procurement of the mating plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The unit uses refrigerant R-449A (A1, H280) with GWP 1397 and 1.35 kg charge, giving 1.9 t CO2 equivalent, which matters for refrigerant handling and leak reporting.

Needed for refrigerant safety classification, environmental reporting and service planning.

Source
“R-449A (A1, H280)”
Installation

The chiller is water-cooled and requires a minimum cooling water differential pressure of 1.5 bar and maximum cooling water pressure of 6 bar.

Ensures the site cooling water supply can meet the hydraulic requirements.

Source
“min. cooling water differential pressure”
Operation

The unit is IP20 and must be operated within an ambient temperature range of 5°C to 40°C.

Defines the environmental limits for installation location and enclosure protection.

Source
“min. ambient temperature”
Installation

The unit is intended for externally closed applications, which constrains how the hydraulic circuit may be connected.

Prevents misapplication in open-loop or unsuitable external circuits.

Source
“For externally closed applications.”
Maintenance

The expansion tank is atmospheric open and the unit has a digital level indicator, so the filling capacity and expansion volume must be respected.

Affects fluid filling, expansion volume planning and maintenance checks.

Source
“Housing, atmospheric open expansion tank and copper soldered evaporator made of stainless steel.”
Performance

Cooling water consumption depends strongly on flow temperature, ranging from 474 l/h at 15°C flow to 264 l/h at -10°C flow.

Needed to size cooling water supply and estimate water consumption.

Source
“Consumption at water 15°C, flow 15°C”
Cumplimiento

The equipment is classified for electromagnetic disturbance as EN55011 Class A, Group 1, which has implications for the installation environment.

Class A equipment may require additional EMC precautions in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”