Selection
The chiller is rated for an operating temperature range of -20 to 100 °C, so it must be matched to the process temperature window.
Defines whether the unit can cover the required process temperatures before selection.
Source
“Operating temperature range
-20...100 °C”
Performance
Cooling capacity drops from 7.5 kW at 20 °C to 2.4 kW at -20 °C, so low-temperature duty needs a larger unit or derating check.
Prevents undersizing when the process runs at the cold end of the range.
Source
“at -20°C
2,4 kW”
Installation
The unit requires a 400 V three-phase supply with 13 A max current and 3x16 A fusing.
Determines electrical supply, cable and protection sizing at the installation site.
Source
“Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
13 A
Fuse (3 phase)
3x16A”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so the site must provide a suitable connection.
Avoids surprises at commissioning because no country plug is supplied.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The pump connection is G1 1/4 male, so hoses and adapters must match this thread.
Ensures correct hydraulic connection and avoids leaks or adapter mismatch.
Source
“Pump connection
G1 1/4 male”
Installation
The minimum filling capacity is 18 l and expansion volume is 48 l, which sets the minimum loop volume and expansion allowance.
Needed to size the external circuit and expansion tank correctly.
Source
“min. filling capacity
18 l
Volume of expansion
48 l”
Installation
The unit is air-cooled and requires ambient temperature between 5 °C and 40 °C.
Room conditions must be checked because cooling capacity falls as ambient rises.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Performance
Cooling capacity is valid at 20 °C room temperature and may drop if ambient temperature rises.
Prevents overestimating capacity in a warm plant room.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Safety
The refrigerant is R-449A with GHS H280, so leak handling and ventilation must account for a pressurised gas hazard.
Affects safety assessment and refrigerant handling procedures.
Source
“R-449A (A1, H280)”
Cumplimiento
The unit is classified Class A, Group 1 for electromagnetic disturbance to EN55011.
Class A equipment may require EMC precautions in residential or sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Voltage and frequency tolerances are ±5 % and ±2 % but not simultaneously in the unfavourable combination.
Site power quality must be checked to avoid invalid supply conditions.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation
The unit is intended for externally closed applications, which constrains the hydraulic circuit design.
Prevents use in open-loop or unsuitable external circuits.
Source
“For externally closed applications.”
Maintenance
The refrigerant charge is 1.03 kg with CO2 equivalent 1.4 t, which matters for leak checks and refrigerant records.
Supports maintenance planning and regulatory refrigerant accounting.
Source
“Refrigerant quantity
1,03 kg
CO2 equivalent
1,4 t”